For the podcast associated with this article, please visit https://academic.oup.com/ehjcr/pages/podcast

An 82-year-old bradycardic and hypotensive woman was admitted after syncope. An electrocardiogram showed third-degree atrioventricular block. The patient reported a history of cardiomegaly. The ultrasound was unavailable. A pacemaker was implanted in the lower portion of the left internal jugular vein with a 6 Fr introducer guided by anatomical landmarks, without any PVCs reported. There was no difficulty in the insertion of the introducer or pacemaker lead. To ensure stimulus capture, a high voltage was maintained. Chest radiography showed the anomalous path of the pacemaker electrode (Figure 1). Chest tomography demonstrated that it was in the mediastinal position (Figure 2). The device was removed since there were no pulmonary, vascular, or pericardial complications. Another pacemaker was successfully placed. The patient did not develop any further symptoms. This case demonstrates the importance of ultrasound or fluoroscopic-guided vascular punctures for pacemaker implantation to avoid injury to the left internal jugular vein, which is easily perforated by a rigid wire or pacemaker lead.1,2

Chest radiography showed the anomalous path of the pacemaker electrode implanted by the left internal jugular vein.
Figure 1

Chest radiography showed the anomalous path of the pacemaker electrode implanted by the left internal jugular vein.

Chest tomography demonstrated pacemaker electrode in the mediastinal position.
Figure 2

Chest tomography demonstrated pacemaker electrode in the mediastinal position.

Consent: The author’s confirm that written consent for submission and publication of this case report including image(s) and associated text has been obtained from the patient’s next of kin in line with COPE guidance.

Conflict of interest: None declared.

Funding: None declared.

References

1

Albertson
TE
,
Fisher
CJ
Jr
,
Vera
Z.
Accidental mediastinal entry via left internal jugular vein annulation
.
Intensive Care Med
1985
;
11
:
154
157
.

2

Lin
J
,
Wang
L
,
Zhao
Y.
Inadvertent left ventricular pacing and perforation by a temporary pacemaker
.
J Electrocardiol
2017
;
50
:
686
689
.

This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided the original work is properly cited. For commercial re-use, please contact [email protected]
Handling Editor: Habib Rehman Khan
Habib Rehman Khan
Handling Editor
Search for other works by this author on:

Editor: David Niederseer,
David Niederseer
Editor
Search for other works by this author on:
Daniel Scherr,
Daniel Scherr
Editor
Search for other works by this author on:
Henning Jansen
Henning Jansen
Editor
Search for other works by this author on:

Comments

0 Comments
Submit a comment
You have entered an invalid code
Thank you for submitting a comment on this article. Your comment will be reviewed and published at the journal's discretion. Please check for further notifications by email.