Abstract

Surgical aortic valve replacement (SAVR) is currently the standard of care to treat patients with severe symptomatic aortic stenosis (AS) and is generally accepted to alleviate symptoms and prolong survival. Based on the results of randomized trials, transcatheter aortic valve implantation (TAVI) is the new standard of care for patients with symptomatic AS who are deemed ‘inoperable’. Debatably, TAVI is also an alternative to SAVR in selected patients who are at high risk but operable. As we approach 10 years of clinical experience with TAVI, with over 50 000 implantations in 40 countries, a review of the current literature and clinical outcomes with this rapidly evolving technology is appropriate.

Introduction

Symptomatic severe aortic stenosis (AS) has a poor prognosis when treated medically and inevitably leads to functional deterioration, heart failure, and death.1 Surgical aortic valve replacement (SAVR) is currently the standard of care and is generally accepted to alleviate symptoms and prolong survival, but ∼30% do not undergo SAVR.2 However, since Dr Alain Cribier pioneered the first transcatheter aortic valve implantation (TAVI) procedure in 2002,3 this relatively new technique has been used extensively in over 40 countries accumulating to >50 000 implantations.4–19 With results from the randomized Placement of AoRTic TraNscathetER Valves (PARTNER) trial,20 TAVI is now the standard of care for extremely high risk or ‘inoperable’ patients and is a valid alternative to surgery for selected high-risk but ‘operable’ patients with symptomatic AS.21 Currently, two different TAVI devices are widely used: the balloon-expandable Edwards SAPIEN Transcatheter Heart Valve (Edwards Lifesciences, Irvine, CA, USA) and the self-expanding Medtronic CoreValve™ (Medtronic, Minneapolis, MN, USA) (Figure 1). Both devices received CE Mark approval for European commercial sale in 2007, and the Edwards SAPIEN valve received FDA pre-market approval in the USA in November 2011. As we approach 10 years of clinical experience with TAVI, a review of the current literature and clinical outcomes is appropriate.

Figure 1

(A) Edwards-SAPIEN Transcatheter Heart Valve (Edwards Lifesciences). (B) Edwards-SAPIEN XT Transcatheter Heart Valve (Edwards Lifesciences). (C) Medtronic CoreValve™ (Medtronic).

Figure 1

(A) Edwards-SAPIEN Transcatheter Heart Valve (Edwards Lifesciences). (B) Edwards-SAPIEN XT Transcatheter Heart Valve (Edwards Lifesciences). (C) Medtronic CoreValve™ (Medtronic).

Initial experience

First-in-man, initial reports, and feasibility studies

Cribier and co-workers3 performed the first TAVI in an inoperable patient in 2002 using a transeptal antegrade approach and a balloon-expandable aortic valve prosthesis, demonstrating the feasibility of percutaneous valve implantation. The antegrade approach was further explored,22 coinciding with the first-in-man retrograde experience of a self-expanding prosthesis (CoreValve™).23

Larger series quickly followed, with early experiences of small initiatives using both the balloon-expandable Cribier valve (Edwards Lifesciences Inc.) and the self-expandable CoreValve™ system.9,19 The devices showed a procedural success rate of ∼80%.

After these single-centre experiences, several larger multicentre feasibility studies were initiated first in Europe and later in the USA.6,24–26 These studies showed that transapical (TA) and transfemoral (TF) TAVI in high-risk patients was feasible and could be performed with a high procedural success rate and a 30-day mortality of ∼10–15% (Table 1).

Table 1

Multicentre feasibility studies

Study Enrollment Number of patients Approach Device Procedural success 30-day mortality 
I-REVIVE/RECAST6 2003–2005 26 Transseptal Edwards SAPIEN 85% (22/26) 16.7% (6/36) 
TF Edwards SAPIEN 57% (4/7) 
Grube et al.9 2005–2007 86 TF CoreValve 74% (64/86) 11.6% (10/86) 
TRAVERCE26 2006–2008 168 TA Edwards SAPIEN 95.8% (161/168) 14.9% (25/168) 
REVIVAL24,25 2006–2008 40 TA Edwards SAPIEN 100% (40/40) 12.5% (7/40) 
2005–2006 55 TF Edwards SAPIEN 87% (48/55) 7.3% (4/55) 
Study Enrollment Number of patients Approach Device Procedural success 30-day mortality 
I-REVIVE/RECAST6 2003–2005 26 Transseptal Edwards SAPIEN 85% (22/26) 16.7% (6/36) 
TF Edwards SAPIEN 57% (4/7) 
Grube et al.9 2005–2007 86 TF CoreValve 74% (64/86) 11.6% (10/86) 
TRAVERCE26 2006–2008 168 TA Edwards SAPIEN 95.8% (161/168) 14.9% (25/168) 
REVIVAL24,25 2006–2008 40 TA Edwards SAPIEN 100% (40/40) 12.5% (7/40) 
2005–2006 55 TF Edwards SAPIEN 87% (48/55) 7.3% (4/55) 

TF, transfemoral; TA, transapical; I-REVIVE, Initial Registry of EndoVascular Implantation of Valves in Europe trial; RECAST, Registry of Endovascular Critical Aortic Stenosis Treatment trial; REVIVAL, PeRcutaneous EndoVascular Implantation of VALves trial; TRAVERCE, The initial multicentre feasibility trial for TA-AVI.

Registries

Edwards registries

Several large European and Canadian registries have been published, showing excellent short- and mid-term results after TAVI using both the TF and TA devices.12,15 The largest registry reported to date is the SOURCE (SAPIEN Aortic Bioprosthesis European Outcome) registry.17,18 Overall, 1038 patients were enrolled at 32 European centres and were treated with either a TF (n = 463) or TA approach (n = 575). Patients treated by TA had more co-morbidities at baseline than TF patients, resulting in a significantly higher EuroSCORE (European System for Cardiac Operative Risk Evaluation) (29.1 vs. 25.7%; P < 0.001). Procedural success was 95.2 and 92.7% and 30-day mortality was 6.3 and 10.3% in the TF and TA populations, respectively. The major limitations of this registry were that >70% of the enrolling centres had no prior experience with TAVI and all adverse events were site-reported without core lab analysis. In early 2011, 1-year results were published, demonstrating a 1-year survival of 76.1% overall, 72.1% for TA and 81.1% for TF patients. Among the surviving patients, 73.5% were New York Class Association (NYHA) class I or II.17

CoreValve™ registries

A number of large dedicated CoreValve registries have been reported; generally, these have been somewhat larger than Edwards registries.14,16 Promising 3-year results were recently reported by Ussia et al.27 and although not yet published, the results of the ADVANCE CoreValve™ registry were presented recently.28 ADVANCE represents a 100% monitored ‘real-world’ experience, with a core laboratory and an independent clinical events committee adjudicating events. The registry included 1015 patients from 44 experienced (>40 prior procedures) centres between March 2010 and July 2011. The mean logistic EuroSCORE was 19.2%. At 30 days and 6 months, the rate of all-cause mortality was 4.5 and 12.8%, respectively, with cardiac mortality of 3.4 and 8.4%, respectively. ADVANCE provides insights into contemporary TAVI data of experienced operators, and is a benchmark for comparing outcomes.

Mixed national registries

In 2011, results from four mixed CoreValve™ and Edwards European national registries have been reported, mostly using the TF and TA routes (Table 2).4,8,29,30 Overall, patients included in these registries were at high-risk according to surgical risk models; mean EuroSCORE 18–30%. These registries showed 1-year survival rates ranging between 71.9 and 81.6%. The UK registry reported the longest follow-up; survival was 73.7% at 2 years.30 Several of these national initiatives performed access-route comparisons and reported that survival was generally higher in patients treated through the TF route.4,30 However, it should be noted that a transfemoral-first approach is often advocated, which may introduce selection bias and an unfair comparison between the two access routes.31

Table 2

Clinical outcomes after TAVI according to access site and device type: major published data

Authors Type of study Number of patients STS (%) Logistic EuroScore (%) Follow-up (months) Procedural success rate (%) Mortality 30-day (%) Mortality 1-year (%) Major access complications 30-day (%) Stroke 30-day (%) Need for new PPM (%) 
Edwards SAPIEN: TF 
 Lefevre et al.12 Registry 61 11.3 25.7 12 95.4 8.2 21.3 16.4 3.3 1.8 
 Eltchaninoff et al.8 Registry 95 17.4 25.6 98.3a 8.4 – 6.3 4.2 5.3 
 Himbert et al.11 Registry 51 15.0 25.0 12 90.0 8.0b 19.0 12.0 6.0 6.0 
 Rodes-Cabau et al.15 Registry 162 9.0 – 24 90.5 9.5 25.0 13.1 3.0 3.6 
 Thomas et al.17,18 Registry 463 – 14.5 95.2 6.3 18.9 22.9 2.4 6.7 
 Leon et al.20 RCT 179 11.2 26.4 12 – 5.0c 30.7c 16.2 6.7d 3.4 
 Bosmans et al.4 Registry 99 – 29.0 12 97.0 6.0 18.0 – 2.0 4.0 
 Smith et al.21 RCT 244 11.7 29.1 12 – 3.3c 22.2c 14.0 3.7d 3.7 
Edwards SAPIEN: TA 
 Walther et al.26 Feasibility study 168 – 27.0 12 95.8 15.0 37.0 1.2 2.0 2.3 
 Svensson et al.25 Feasibility study 40 13.4 35.5 87.5 17.5 – – 5.0 – 
 Lefevre et al.12 Registry 69 11.3 33.8 12 96.4 18.8 50.7 5.8e 1.5 3.8 
 Eltchaninoff et al.8 Registry 71 18.4 26.8 98.3a 16.9 – 5.6f 2.8 5.6 
 Himbert et al.11 Registry 24 18.0 28.0 12 100 16.0b 26.0 8.0 4.0 
 Rodes-Cabau et al.15 Registry 177 10.5 – 96.1 11.3 22.0 13.0f 1.7 6.2 
 Thomas et al.17,18 Registry 575 – 16.3 95.7 10.3 27.9 4.7 2.6 7.3 
 Bosmans et al.4 Registry 88 – 33.0 12 97.0 14.0 37.0 – 8.0 6.0 
 Smith et al.21 RCT 104 11.8 29.8 12 – 3.8c 29.0c 3.8 6.8 3.9 
 D'Onofrio et al.7 Registry 504 11.0 26.3 24 99.0 8.3 18.8 – 3.0 5.4 
Medtronic CoreValve™: TF 
 Tamburino et al.16 Registry 663 – 23.0 12 98.0 5.4 15.0 2.0 2.5g 17.4 
 Bosmans et al.4 Registry 133 – 25.0 12 98.0 11.0 22.0 – 4.0 22.0 
 Grube et al.10 Registry 86 – 21.6 88.0 12.0 – – 10.0 – 
 Piazza et al.14 Registry 646 – 23.1 97.2 8.0 – 1.9 1.9 9.3 
 Eltchaninoff et al.8 Registry 66 21.3 24.7 98.3a 15.1 – 7.5 4.5 25.7 
 Petronio et al.13 Registry 460 – 19.4 98.4 6.1 11.4 2.0 1.7 16.1 
 Buellesfeld et al.5 Registry 126ξ h – 23.0 24 72.6 15.2 28.1 – 9.6 26.2 
Medtronic CoreValve™: SC 
 Eltchaninoff et al.8 Registry 12 21.0 24.6 98.3a 8.3 – 8.3 25.0 
 Bosmans et al.4 Registry – 25.0 12 98.0 11.0 – 4.0 22.0 
 Petronio et al.13 Registry 54 – 25.3 100 6.7 1.9 18.5 
 Zahn et al.29i Registry 697 – 20.5 98.4 12.4 – 19.5 2.8 39.3j 
Authors Type of study Number of patients STS (%) Logistic EuroScore (%) Follow-up (months) Procedural success rate (%) Mortality 30-day (%) Mortality 1-year (%) Major access complications 30-day (%) Stroke 30-day (%) Need for new PPM (%) 
Edwards SAPIEN: TF 
 Lefevre et al.12 Registry 61 11.3 25.7 12 95.4 8.2 21.3 16.4 3.3 1.8 
 Eltchaninoff et al.8 Registry 95 17.4 25.6 98.3a 8.4 – 6.3 4.2 5.3 
 Himbert et al.11 Registry 51 15.0 25.0 12 90.0 8.0b 19.0 12.0 6.0 6.0 
 Rodes-Cabau et al.15 Registry 162 9.0 – 24 90.5 9.5 25.0 13.1 3.0 3.6 
 Thomas et al.17,18 Registry 463 – 14.5 95.2 6.3 18.9 22.9 2.4 6.7 
 Leon et al.20 RCT 179 11.2 26.4 12 – 5.0c 30.7c 16.2 6.7d 3.4 
 Bosmans et al.4 Registry 99 – 29.0 12 97.0 6.0 18.0 – 2.0 4.0 
 Smith et al.21 RCT 244 11.7 29.1 12 – 3.3c 22.2c 14.0 3.7d 3.7 
Edwards SAPIEN: TA 
 Walther et al.26 Feasibility study 168 – 27.0 12 95.8 15.0 37.0 1.2 2.0 2.3 
 Svensson et al.25 Feasibility study 40 13.4 35.5 87.5 17.5 – – 5.0 – 
 Lefevre et al.12 Registry 69 11.3 33.8 12 96.4 18.8 50.7 5.8e 1.5 3.8 
 Eltchaninoff et al.8 Registry 71 18.4 26.8 98.3a 16.9 – 5.6f 2.8 5.6 
 Himbert et al.11 Registry 24 18.0 28.0 12 100 16.0b 26.0 8.0 4.0 
 Rodes-Cabau et al.15 Registry 177 10.5 – 96.1 11.3 22.0 13.0f 1.7 6.2 
 Thomas et al.17,18 Registry 575 – 16.3 95.7 10.3 27.9 4.7 2.6 7.3 
 Bosmans et al.4 Registry 88 – 33.0 12 97.0 14.0 37.0 – 8.0 6.0 
 Smith et al.21 RCT 104 11.8 29.8 12 – 3.8c 29.0c 3.8 6.8 3.9 
 D'Onofrio et al.7 Registry 504 11.0 26.3 24 99.0 8.3 18.8 – 3.0 5.4 
Medtronic CoreValve™: TF 
 Tamburino et al.16 Registry 663 – 23.0 12 98.0 5.4 15.0 2.0 2.5g 17.4 
 Bosmans et al.4 Registry 133 – 25.0 12 98.0 11.0 22.0 – 4.0 22.0 
 Grube et al.10 Registry 86 – 21.6 88.0 12.0 – – 10.0 – 
 Piazza et al.14 Registry 646 – 23.1 97.2 8.0 – 1.9 1.9 9.3 
 Eltchaninoff et al.8 Registry 66 21.3 24.7 98.3a 15.1 – 7.5 4.5 25.7 
 Petronio et al.13 Registry 460 – 19.4 98.4 6.1 11.4 2.0 1.7 16.1 
 Buellesfeld et al.5 Registry 126ξ h – 23.0 24 72.6 15.2 28.1 – 9.6 26.2 
Medtronic CoreValve™: SC 
 Eltchaninoff et al.8 Registry 12 21.0 24.6 98.3a 8.3 – 8.3 25.0 
 Bosmans et al.4 Registry – 25.0 12 98.0 11.0 – 4.0 22.0 
 Petronio et al.13 Registry 54 – 25.3 100 6.7 1.9 18.5 
 Zahn et al.29i Registry 697 – 20.5 98.4 12.4 – 19.5 2.8 39.3j 

TAVI, transcatheter aortic valve implantation; TF, transfemoral; TA, transapical; SC, subclavian; STS, Society of Thoracic Surgeons; EuroSCORE, European System for Cardiac Operative Risk Evaluation; RCT, randomized controlled trial; PPM, permanent pacemaker.

aGlobal procedural success rate including Edwards SAPIEN TF, Edwards SAPIEN TA and Medtronic CoreValve TF and subclavian was 98.3%.

bIn-hospital mortality.

cIntention-to-treat mortality rate.

dMajor and minor stroke.

eVascular-related complications.

fApex-related complications.

g1-year.

h124 patients received a TF approach and 2 an SC approach.

iOutcomes reported together: 566 (81.2%) Medtronic CoreValve™ TF, 22 (3.2%) Medtronic CoreValve™ SC, 106 (15.2%) Edwards SAPIEN TF.

jMedtronic CoreValve™ 42.5%, Edwards SAPIEN 22%.

Recently, the largest registry to date was reported by the FRANCE 2 (FRench Aortic National CoreValve and Edwards) investigators.32 They included 3195 patients treated between January 2010 and December 2011 at 34 centres. The registry reflects contemporary real-life use of available TAVI devices in patients at high surgical risk; the Edwards SAPIEN and the Medtronic CoreValve devices were used in, respectively, 66.9 and 33.1%. The transfemoral approach was most popular (74.6%), followed by transapical (17.8%) and subclavian (5.8%), while 1.8% underwent some other approach. The procedural success rate was 96.9% and 1-year survival in patients was 76.0%.

Randomized trials

Completed trials

While registry reports are of crucial value to assess ‘real-world’ use of TAVI, more rigorous assessments are available from the first multicentre, randomized clinical PARTNER trials (Placement of Aortic Transcatheter Valves; ClinicalTrials.gov Identifier: NCT00530894) (Figure 2).20,21

Figure 2

PARTNER trial I design.

Figure 2

PARTNER trial I design.

As the first of two parallel trials was completed, the results of PARTNER IB showed that TF TAVI was superior to standard therapy in patients not deemed candidates for surgery.20 The primary endpoint of all-cause mortality was markedly reduced by 46% (P < 0.001). Recently reported 2-year outcomes showed continued encouraging results (Figure 3A).33 At 2 years, the primary endpoint of all-cause mortality was reduced from 67.6% in the standard treatment arm to 43.3% in the TAVI arm (P < 0.001).

Figure 3

(A) Two-year with 1-year landmark analysis of all-cause mortality Kaplan–Meier curve in PARTNER trial cohort 1B. Reprinted with permission from Leon and colleagues20 and Makkar and colleagues.33 (B) Two-year all cause-mortality Kaplan–Meier curve in PARTNER trial cohort 1A. Adapted with permission from Kodali and colleagues.34 (C) Two-year stroke Kaplan–Meier curve in PARTNER trial cohort 1A. Adapted with permission from Kodali and colleagues.34

Figure 3

(A) Two-year with 1-year landmark analysis of all-cause mortality Kaplan–Meier curve in PARTNER trial cohort 1B. Reprinted with permission from Leon and colleagues20 and Makkar and colleagues.33 (B) Two-year all cause-mortality Kaplan–Meier curve in PARTNER trial cohort 1A. Adapted with permission from Kodali and colleagues.34 (C) Two-year stroke Kaplan–Meier curve in PARTNER trial cohort 1A. Adapted with permission from Kodali and colleagues.34

The PARTNER cohort IA compared TAVI with SAVR and met its non-inferiority endpoint: the all-cause 1-year mortality in the TAVI group was non-inferior to the SAVR group (24.2 vs. 26.8%; P = 0.44; P = 0.001 for non-inferiority).21 Some concerns were raised with regard to neurologic events that were somewhat higher with TAVI than SAVR at 30 days (5.5 vs. 2.4%; P = 0.04) and 1 year (8.3 vs. 4.3%; P = 0.04). Although the recently published 2-year results showed that stroke rates were similar for TAVI and SAVR during 1 and 2 years with a hazard ratio of 1.22 (95% CI 0.67–2.23, P = 0.52), the issue of stroke warrants further investigation and should not be underestimated (Figure 3B and C).34 The rate of the composite of all-cause death and stroke was encouragingly nearly identical after TAVI (37.1%) and SAVR (36.4%) at 2 years (P = 0.85).

Ongoing trials

In the USA, a randomized trial is currently ongoing to evaluate the safety and efficacy of the Medtronic CoreValve™ in the treatment of severe symptomatic AS in patients at high or extreme risk for SAVR (ClinicalTrials.gov Identifier: NCT01240902). The trial consists of two arms. Patients in a high-risk arm will be randomized between SAVR and TAVI; the primary endpoint consists of all-cause mortality at 1 year. An extreme risk arm will function as an observational arm in which a composite of all-cause mortality and major stroke is the primary endpoint.

As a sequel to the PARTNER I trial, a second randomized trial (PARTNER II) is currently ongoing. It was designed to investigate the performance and outcomes after TAVI with the next-generation Edwards SAPIEN XT valve, model 9300TFX, as well as the new low-profile 18-Fr NovaFlex™ delivery catheter in patients deemed non-operable (ClinicalTrials.gov Identifier: NCT01314313) (Figure 4A). Given the results of the control arm in PARTNER IB, it has been judged that a study comparing TAVI against a ‘medical management’ control group is no longer ethical.35 Consequently, an ‘old device’ vs. ‘new device’ non-inferiority trial was designed. Enrolment began in January 2011 and it is anticipated that primary endpoint results will be published mid-2013.

Figure 4

(A) PARTNER trial 2 design. (B) SURTAVI trial design (provided by Medtronic on 26 April 2012).

Figure 4

(A) PARTNER trial 2 design. (B) SURTAVI trial design (provided by Medtronic on 26 April 2012).

In Denmark, a phase 2 randomized trial evaluating TAVI in patients ≥70 years of age started enrolment in December 2009 (ClinicalTrials.goc identifier: NCT01057173). The trial will randomize a total of 280 patients to TAVI (n = 140) and SAVR (n = 140). The primary endpoint is the composite of all-cause death, myocardial infarction, and stroke at 1 year and is scheduled to be completed late 2013.

In an attempt to expand the indication of TAVI to lower-risk patients, the PARTNER IIA trial will be randomizing patients between TAVI with the SAPIEN XT valve and SAVR in intermediate risk patients (ClinicalTrials.gov Identifier: NCT01314313) (Figure 4A). Similarly, the prospective randomized, international SURTAVI trial will randomize 1900 intermediate risk patients between TAVI with the Medtronic CoreValve™ and SAVR at ∼80 centres throughout the USA, Canada, Europe, and Australia (ClinicalTrials.gov Identifier: NCT01586910) (Figure 4B).

Cost-effectiveness

Since TAVI has been shown to be superior to standard medical therapy and non-inferior to SAVR and is increasingly being used in current practice, the incremental costs and cost-effectiveness of this therapy warrant evaluation.

In the PARTNER IB trial, the mean cost for TF TAVI was $42 806 which accumulated to $78 542 for the initial hospitalization and $106 076 at 1 year.36 Compared with medical therapy, TAVI was ∼$52 455 (95% CI, $40 635–$64 275) more expensive at 1 year, but quality of life was significantly better in patients who underwent TAVI. This resulted in an incremental cost-effectiveness ratio of $50 212 per life-year gained, and $61 889 per quality-adjusted life-year (QALY). The authors rightfully concluded that for patients not candidates for surgery in the USA, TAVI increases (quality-adjusted) life-years at reasonable costs similar to other cardiovascular technologies.

In the PARTNER IA trial, similar costs were found in TF patients as compared with the PARTNER IB trial; $71 955 for the index hospitalization and $94 206 at 1 year, which was comparable to patients who underwent SAVR ($74 452 and $96 417, respectively). However, there was only a minor gain in the number of life-years (0.065: 95% CI, 0.011–0.125) and QALYs (0.068: 95% CI, 0.017–0.123) in comparison with SAVR. Through bootstrap analysis it was concluded that TF TAVI cost was <$50 000 per QALY in 74.7% of times, clearly demonstrating cost-effectiveness in the USA. Patients who could not undergo TF due to anticipated vascular and/or bleeding complications were randomized between TA TAVI (n = 101) and SAVR (n = 91). The index hospitalization was more expensive in the TA group, although not significantly so ($90 548 vs. $79 540, P = 0.08). At 1-year follow-up, costs accumulated to a mean of $107 779 for TA and $98 183 for SAVR, with a small detriment in life-years (−0.015: 95% CI, −0.103–0.080) and QALYs (−0.070: 95% CI, −0.151–0.012). Therefore, TA TAVI was found to be a less attractive alternative to SAVR, although this conclusion has been somewhat criticized because the analysis was not powered and operators were little experienced.31

Alternative access sites

Like the TA approach, a subclavian approach allows patients with unfavourable iliofemoral anatomy or extensive disease to be treated with TAVI. Petronio et al.13 recently reported a series of 54 patients, showing a procedural success rate of 100%, a procedural mortality of 0, a 30-day mortality of 0%, and 6-month mortality of 9.4%. No specific vascular complications for subclavian access were reported. The subclavian approach is usually performed with the self-expanding CoreValve™ system and can be fully percutaneous.37

Recently, a transaortic approach with direct access to the ascending aorta though an anterior minithoracotomy has been advocated. Access is gained through a J-shaped partial upper sternotomy or using a small right thoracotomy through the intercostal space. Avoidance of LV apical injury or inadequate healing along with reduction in post-operative pain and its associated impairment of respiratory dynamics are potential advantages of this novel approach. Encouraging results have been published from small series using both devices.38,39 It may be suitable for patients with unfavourable iliofemoral and subclavian anatomy and in whom a TA approach is considered too risky (chest deformity, severe respiratory disease or low ejection fraction). Also, TAVI via the carotid artery has been proposed. In such cases, it is crucial to evaluate the cerebral arteries, carotid and vertebral arteries, and circle of Willis, to assess the risk of ischaemic stroke.40

Valve-in-valve for failing bioprostheses

Since 2007, when the first TAVI was implanted in a failing surgical aortic bioprosthesis in order to avoid redo surgery, interest in this concept has grown and feasibility and safety have been established.41,42 Piazza and colleagues43 published a series of 20 patients (mostly TA: 16/20) and reported successful implantation in 18 of 20 patients and in-hospital mortality in 3 patients. Indeed, transcatheter heart valves have also been implanted in failing mitral prostheses or even annuloplasty rings, and failing tricuspid prostheses, expanding the potential use of devices originally developed for the aortic position.44,45

Knowledge of the basic construction, dimensions, and potential failure modes of the surgical bioprostheses is of paramount importance for this technique to succeed. Various complications such as coronary obstruction and device embolization may be implicated with certain surgical bioprostheses but not others.46 Also, small surgical bioprostheses (e.g. 19 mm) may not respond well to valve-in-valve implantation because of device constraint within the rigid bioprosthesis and incomplete stent expansion, frequently leading to prosthesis–patient mismatch.47,48

The presence of a functioning mitral prosthesis may further complicate device delivery, although a recent report has shown that optimal valve positioning through a TA approach should be technically achievable with modifications of the ‘classic’ procedure.49

Efficacy and long-term outcomes

Symptom improvement

Improvement in cardiac symptoms and functional class has been reported at short- and medium-term after TAVI.14,18,20,21,50 However, functional assessment of the population currently eligible for and treated with TAVI is difficult, mainly because of their multiple co-morbidities.51

Three-year follow-up data have been published and are consistent with lasting improvement in cardiac symptoms.52 While 86% of patients were in NYHA class III or IV at baseline, 93% of surviving patients were in NYHA class I/II at 3-year follow-up. Similarly, the PARTNER trial showed that patients treated with TAVI compared with patients treated with standard medical therapy have better symptom control at 1 year.20 Indeed, the 1-year rate of NYHA class III or IV was 25.2% for the TAVI group compared with 58.0% for the standard medical therapy group (P < 0.001).

Valve durability and haemodynamic performance

TAVI has demonstrated excellent immediate and short-term durability of the prosthesis that is comparable to SAVR, sustaining to 3 years.14,15,18,20,27,34,52 Actually, data suggest that transcatheter heart valves have greater valve areas and lower gradients than surgical bioprostheses (Figure 5),34,53 which could reduce the prevalence of prosthesis–patient mismatch.54 For both the Edwards SAPIEN and Medtronic CoreValve™ there was no evidence of structural or non-structural valvular deterioration, stent fracture, deformation, or valve migration.

Figure 5

Two-year time trends in haemodynamics after TAVI vs. SAVR. Adapted with permission from Kodali and colleagues.34

Figure 5

Two-year time trends in haemodynamics after TAVI vs. SAVR. Adapted with permission from Kodali and colleagues.34

Predictors

As emphasized throughout the manuscript, many of the listed complications are predictors of short-term and/or long-term mortality. As current randomized trials are moving towards evaluating TAVI in a lower-risk patient population (SURTAVI, PARTNER 2) with a longer life expectancy, prediction of mid- and long-term outcomes (≥1 year) will become increasingly important. Some predictors should be similar to the surgical literature, but the mounting TAVI experience has shown that the incidence and ratio may vary significantly between the two therapies. For example, paravalvular leakage is more common after TAVI than after SAVR and has been identified as a potential significant predictor for long-term mortality.34

Table 3 provides a summary of independent predictors of mortality that have been identified in previous studies. Due to the relative infancy of TAVI and the lack of large databases for SAVR,55 it is likely that additional predictors will come to light over the years. Furthermore, accurate hazard ratios of predictors cannot be given at the current time, due to the severe heterogeneity between studies.

Table 3

Independent predictors of long-term mortality after TAVI

Advanced age 
Smoking 
Logistic EuroSCORE 
STS score 
Calcium score 
Baseline renal failure 
Baseline anaemia 
Pulmonary hypertension 
COPD 
Liver disease 
Prior stroke 
Post-procedural PVL ≥2+ 
Myocardial injury 
Systematic inflammatory response syndrome 
Major vascular complication 
Acute kidney injury 
Early experience with TAVI 
Advanced age 
Smoking 
Logistic EuroSCORE 
STS score 
Calcium score 
Baseline renal failure 
Baseline anaemia 
Pulmonary hypertension 
COPD 
Liver disease 
Prior stroke 
Post-procedural PVL ≥2+ 
Myocardial injury 
Systematic inflammatory response syndrome 
Major vascular complication 
Acute kidney injury 
Early experience with TAVI 

TAVI, transcatheter aortic valve implantation.

Lessons learned

Patient selection

One of the critical aspects of TAVI we have learned so far is that patient selection is crucial but cumbersome due to inaccuracy of current risk models to predict outcomes in high-risk patients.56 Several variables that have shown to be predictive are not included, such as frailty, liver disease, and the presence of a porcelain aorta. Decision making should therefore not be based exclusively on clinical risk scores. Instead, it is accepted that the heart team can better judge patient eligibility for TAVI or SAVR. Such a team is dynamic and can include general cardiologists, interventional cardiologists, surgeons, imaging specialists, neurologists, anaesthesiologists, geriatricians, and other specialists.57,58

Besides the decision to treat by means of TAVI or SAVR, one must consider multiple access approaches. Frequently, a ‘transfemoral-first’ attitude is advocated and comprehensive screening of the peripheral arteries and aorta by angiography or preferably by multislice CT-scan (MSCT), is necessary to assess feasibility.59 MSCT also allows for evaluation of left ventricular dimensions and function, and other potential diseases (e.g. coronary artery disease), which can further help to contemplate feasibility, safety, and efficacy.

Sizing

Accurate preoperative annular sizing is one of the main predictors of a successful TAVI procedure. Several modalities have been proposed for accurate sizing. At first, trans-thoracic and/or trans-oesophageal echocardiography were used to decide which size valve would best be implanted to achieve procedural success with limited or no residual para-valvular aortic regurgitation. More recently, the use of three-dimensional and even four-dimensional MSCT has been shown to be most effective in sizing for TAVI.60,61 In contrast to trans-esophageal echocardiography, it is non-invasive and has a high reproducibility.62 Recent studies have shown that the area-derived diameter and basal ring average diameter of the annulus are the most suitable measurements for valve-sizing. Nevertheless, oversizing of the transcatheter heart valve in relation to the annulus size remains necessary to obtain procedural success with limited aortic regurgitation.

Learning curve

Understanding the importance of patient selection, utilizing better anatomical screening to clarify both the aortic root and iliofemoral geometry, and the development of new devices have led to notable improvements in outcome over time. A report highlighting the importance of the learning curve in 270 patients showed that procedural experience was an independent predictor of 30-day survival.63 Furthermore, the procedural success rate has significantly increased (Figure 6);63,64 the use of contrast volume use and radiation doses has decreased;65 and procedural complications have declined.66

Figure 6

Effect of the learning curve on procedural success (A) and 30-day mortality (B) after TAVI. Data from Gurvitch and colleagues.63 (C) Single centre experience on 136 patients comparing three generations of CoreValve™ devices. Data from Grube and colleagues.64

Figure 6

Effect of the learning curve on procedural success (A) and 30-day mortality (B) after TAVI. Data from Gurvitch and colleagues.63 (C) Single centre experience on 136 patients comparing three generations of CoreValve™ devices. Data from Grube and colleagues.64

Conflict of interest: P.G. has received consulting fees/honoraria from Edwards Lifesciences. D.W. and J.W. have received consulting fees from Edwards Lifesciences and St Jude Medical. M.R.W. and S.K. have received consulting fees from Edwards Lifesciences. A.P.K. is a member of the steering committee for the SURTAVI trial. M.B.L. is a nonpaid member of the Scientific Advisory Boards of Edwards Lifesciences and Medtronic Vascular. The other authors report no conflicts of interest.

References

1
Ross
J
Jr
Braunwald
E
Aortic stenosis
Circulation
 , 
1968
, vol. 
38
 (pg. 
61
-
67
)
2
Bach
DS
Siao
D
Girard
SE
Duvernoy
C
McCallister
BD
Jr
Gualano
SK
Evaluation of patients with severe symptomatic aortic stenosis who do not undergo aortic valve replacement: the potential role of subjectively overestimated operative risk
Circ Cardiovasc Qual Outcomes
 , 
2009
, vol. 
2
 (pg. 
533
-
539
)
3
Cribier
A
Eltchaninoff
H
Bash
A
Borenstein
N
Tron
C
Bauer
F
Derumeaux
G
Anselme
F
Laborde
F
Leon
MB
Percutaneous transcatheter implantation of an aortic valve prosthesis for calcific aortic stenosis: first human case description
Circulation
 , 
2002
, vol. 
106
 (pg. 
3006
-
3008
)
4
Bosmans
JM
Kefer
J
De Bruyne
B
Herijgers
P
Dubois
C
Legrand
V
Verheye
S
Rodrigus
I
Belgian TAVI Registry Participants
Procedural, 30-day and one year outcome following CoreValve or Edwards transcatheter aortic valve implantation: results of the Belgian national registry
Interact Cardiovasc Thorac Surg
 , 
2011
, vol. 
12
 (pg. 
762
-
767
)
5
Buellesfeld
L
Gerckens
U
Schuler
G
Bonan
R
Kovac
J
Serruys
PW
Labinaz
M
den Heijer
P
Mullen
M
Tymchak
W
Windecker
S
Mueller
R
Grube
E
2-year follow-up of patients undergoing transcatheter aortic valve implantation using a self-expanding valve prosthesis
J Am Coll Cardiol
 , 
2011
, vol. 
57
 (pg. 
1650
-
1657
)
6
Cribier
A
Eltchaninoff
H
Tron
C
Bauer
F
Agatiello
C
Nercolini
D
Tapiero
S
Litzler
PY
Bessou
JP
Babaliaros
V
Treatment of calcific aortic stenosis with the percutaneous heart valve: mid-term follow-up from the initial feasibility studies: the French experience
J Am Coll Cardiol
 , 
2006
, vol. 
47
 (pg. 
1214
-
1223
)
7
D'Onofrio
A
Rubino
P
Fusari
M
Salvador
L
Musumeci
F
Rinaldi
M
Vitali
EO
Glauber
M
Di Bartolomeo
R
Alfieri
OR
Polesel
E
Aiello
M
Casabona
R
Livi
U
Grossi
C
Cassese
M
Pappalardo
A
Gherli
T
Stefanelli
G
Faggian
GG
Gerosa
G
Clinical and hemodynamic outcomes of ‘all-comers’ undergoing transapical aortic valve implantation: results from the Italian Registry of Trans-Apical Aortic Valve Implantation (I-TA)
J Thorac Cardiovasc Surg
 , 
2011
, vol. 
142
 (pg. 
768
-
775
)
8
Eltchaninoff
H
Prat
A
Gilard
M
Leguerrier
A
Blanchard
D
Fournial
G
Iung
B
Donzeau-Gouge
P
Tribouilloy
C
Debrux
JL
Pavie
A
Gueret
P
FRANCE Registry Investigators
Transcatheter aortic valve implantation: early results of the FRANCE (FRench Aortic National CoreValve and Edwards) registry
Eur Heart J
 , 
2011
, vol. 
32
 (pg. 
191
-
197
)
9
Grube
E
Laborde
JC
Gerckens
U
Felderhoff
T
Sauren
B
Buellesfeld
L
Mueller
R
Menichelli
M
Schmidt
T
Zickmann
B
Iversen
S
Stone
GW
Percutaneous implantation of the CoreValve self-expanding valve prosthesis in high-risk patients with aortic valve disease: the Siegburg first-in-man study
Circulation
 , 
2006
, vol. 
114
 (pg. 
1616
-
1624
)
10
Grube
E
Schuler
G
Buellesfeld
L
Gerckens
U
Linke
A
Wenaweser
P
Sauren
B
Mohr
FW
Walther
T
Zickmann
B
Iversen
S
Felderhoff
T
Cartier
R
Bonan
R
Percutaneous aortic valve replacement for severe aortic stenosis in high-risk patients using the second- and current third-generation self-expanding CoreValve prosthesis: device success and 30-day clinical outcome
J Am Coll Cardiol
 , 
2007
, vol. 
50
 (pg. 
69
-
76
)
11
Himbert
D
Descoutures
F
Al-Attar
N
Iung
B
Ducrocq
G
Detaint
D
Brochet
E
Messika-Zeitoun
D
Francis
F
Ibrahim
H
Nataf
P
Vahanian
A
Results of transfemoral or transapical aortic valve implantation following a uniform assessment in high-risk patients with aortic stenosis
J Am Coll Cardiol
 , 
2009
, vol. 
54
 (pg. 
303
-
311
)
12
Lefevre
T
Kappetein
AP
Wolner
E
Nataf
P
Thomas
M
Schachinger
V
De Bruyne
B
Eltchaninoff
H
Thielmann
M
Himbert
D
Romano
M
Serruys
P
Wimmer-Greinecker
G
PARTNER EU Investigator Group
One year follow-up of the multi-centre European PARTNER transcatheter heart valve study
Eur Heart J
 , 
2011
, vol. 
32
 (pg. 
148
-
157
)
13
Petronio
AS
De Carlo
M
Bedogni
F
Marzocchi
A
Klugmann
S
Maisano
F
Ramondo
A
Ussia
GP
Ettori
F
Poli
A
Brambilla
N
Saia
F
De Marco
F
Colombo
A
Safety and efficacy of the subclavian approach for transcatheter aortic valve implantation with the CoreValve revalving system
Circ Cardiovasc Interv
 , 
2010
, vol. 
3
 (pg. 
359
-
366
)
14
Piazza
N
Grube
E
Gerckens
U
den Heijer
P
Linke
A
Luha
O
Ramondo
A
Ussia
G
Wenaweser
P
Windecker
S
Laborde
JC
de Jaegere
P
Serruys
PW
Procedural and 30-day outcomes following transcatheter aortic valve implantation using the third generation (18 Fr) corevalve revalving system: results from the multicentre, expanded evaluation registry 1-year following CE mark approval
EuroIntervention
 , 
2008
, vol. 
4
 (pg. 
242
-
249
)
15
Rodes-Cabau
J
Webb
JG
Cheung
A
Ye
J
Dumont
E
Feindel
CM
Osten
M
Natarajan
MK
Velianou
JL
Martucci
G
DeVarennes
B
Chisholm
R
Peterson
MD
Lichtenstein
SV
Nietlispach
F
Doyle
D
DeLarochelliere
R
Teoh
K
Chu
V
Dancea
A
Lachapelle
K
Cheema
A
Latter
D
Horlick
E
Transcatheter aortic valve implantation for the treatment of severe symptomatic aortic stenosis in patients at very high or prohibitive surgical risk: acute and late outcomes of the multicenter Canadian experience
J Am Coll Cardiol
 , 
2010
, vol. 
55
 (pg. 
1080
-
1090
)
16
Tamburino
C
Capodanno
D
Ramondo
A
Petronio
AS
Ettori
F
Santoro
G
Klugmann
S
Bedogni
F
Maisano
F
Marzocchi
A
Poli
A
Antoniucci
D
Napodano
M
De Carlo
M
Fiorina
C
Ussia
GP
Incidence and predictors of early and late mortality after transcatheter aortic valve implantation in 663 patients with severe aortic stenosis
Circulation
 , 
2011
, vol. 
123
 (pg. 
299
-
308
)
17
Thomas
M
Schymik
G
Walther
T
Himbert
D
Lefevre
T
Treede
H
Eggebrecht
H
Rubino
P
Colombo
A
Lange
R
Schwarz
RR
Wendler
O
One-year outcomes of cohort 1 in the Edwards SAPIEN Aortic Bioprosthesis European Outcome (SOURCE) registry: the European registry of transcatheter aortic valve implantation using the Edwards SAPIEN valve
Circulation
 , 
2011
, vol. 
124
 (pg. 
425
-
433
)
18
Thomas
M
Schymik
G
Walther
T
Himbert
D
Lefevre
T
Treede
H
Eggebrecht
H
Rubino
P
Michev
I
Lange
R
Anderson
WN
Wendler
O
Thirty-day results of the SAPIEN aortic Bioprosthesis European Outcome (SOURCE) Registry: A European registry of transcatheter aortic valve implantation using the Edwards SAPIEN valve
Circulation
 , 
2010
, vol. 
122
 (pg. 
62
-
69
)
19
Webb
JG
Chandavimol
M
Thompson
CR
Ricci
DR
Carere
RG
Munt
BI
Buller
CE
Pasupati
S
Lichtenstein
S
Percutaneous aortic valve implantation retrograde from the femoral artery
Circulation
 , 
2006
, vol. 
113
 (pg. 
842
-
850
)
20
Leon
MB
Smith
CR
Mack
M
Miller
DC
Moses
JW
Svensson
LG
Tuzcu
EM
Webb
JG
Fontana
GP
Makkar
RR
Brown
DL
Block
PC
Guyton
RA
Pichard
AD
Bavaria
JE
Herrmann
HC
Douglas
PS
Petersen
JL
Akin
JJ
Anderson
WN
Wang
D
Pocock
S
PARTNER Trial Investigators
Transcatheter aortic-valve implantation for aortic stenosis in patients who cannot undergo surgery
N Engl J Med
 , 
2010
, vol. 
363
 (pg. 
1597
-
1607
)
21
Smith
CR
Leon
MB
Mack
MJ
Miller
DC
Moses
JW
Svensson
LG
Tuzcu
EM
Webb
JG
Fontana
GP
Makkar
RR
Williams
M
Dewey
T
Kapadia
S
Babaliaros
V
Thourani
VH
Corso
P
Pichard
AD
Bavaria
JE
Herrmann
HC
Akin
JJ
Anderson
WN
Wang
D
Pocock
SJ
PARTNER Trial Investigators
Transcatheter versus surgical aortic-valve replacement in high-risk patients
N Engl J Med
 , 
2011
, vol. 
364
 (pg. 
2187
-
2198
)
22
Cribier
A
Eltchaninoff
H
Tron
C
Bauer
F
Agatiello
C
Sebagh
L
Bash
A
Nusimovici
D
Litzler
PY
Bessou
JP
Leon
MB
Early experience with percutaneous transcatheter implantation of heart valve prosthesis for the treatment of end-stage inoperable patients with calcific aortic stenosis
J Am Coll Cardiol
 , 
2004
, vol. 
43
 (pg. 
698
-
703
)
23
Grube
E
Laborde
JC
Zickmann
B
Gerckens
U
Felderhoff
T
Sauren
B
Bootsveld
A
Buellesfeld
L
Iversen
S
First report on a human percutaneous transluminal implantation of a self-expanding valve prosthesis for interventional treatment of aortic valve stenosis
Catheter Cardiovasc Interv
 , 
2005
, vol. 
66
 (pg. 
465
-
469
)
24
Kodali
SK
O'Neill
WW
Moses
JW
Williams
M
Smith
CR
Tuzcu
M
Svensson
LG
Kapadia
S
Hanzel
G
Kirtane
AJ
Leon
MB
Early and late (one year) outcomes following transcatheter aortic valve implantation in patients with severe aortic stenosis (from the United States REVIVAL trial)
Am J Cardiol
 , 
2011
, vol. 
107
 (pg. 
1058
-
1064
)
25
Svensson
LG
Dewey
T
Kapadia
S
Roselli
EE
Stewart
A
Williams
M
Anderson
WN
Brown
D
Leon
M
Lytle
B
Moses
J
Mack
M
Tuzcu
M
Smith
C
United States feasibility study of transcatheter insertion of a stented aortic valve by the left ventricular apex
Ann Thorac Surg
 , 
2008
, vol. 
86
 (pg. 
46
-
54
discussion 54–5
26
Walther
T
Kasimir
MT
Doss
M
Schuler
G
Simon
P
Schachinger
V
Mohr
FW
Wimmer-Greinecker
G
One-year interim follow-up results of the TRAVERCE trial: the initial feasibility study for trans-apical aortic-valve implantation
Eur J Cardiothorac Surg
 , 
2011
, vol. 
39
 (pg. 
532
-
537
)
27
Ussia
GP
Barbanti
M
Petronio
AS
Tarantini
G
Ettori
F
Colombo
A
Violini
R
Ramondo
A
Santoro
G
Klugmann
S
Bedogni
F
Maisano
F
Marzocchi
A
Poli
A
De Carlo
M
Napodano
M
Fiorina
C
De Marco
F
Antoniucci
D
de Cillis
E
Capodanno
D
Tamburino
C
on behalf of CoreValve Italian Registry Investigators
Transcatheter aortic valve implantation: 3-year outcomes of self-expanding CoreValve prosthesis
Eur Heart J
 , 
2012
, vol. 
33
 (pg. 
969
-
976
)
28
Wood
S
TAVI registry updates ADVANCE the field, raise hopes for lower stroke rates
 
29
Zahn
R
Gerckens
U
Grube
E
Linke
A
Sievert
H
Eggebrecht
H
Hambrecht
R
Sack
S
Hauptmann
KE
Richardt
G
Figulla
HR
Senges
J
German Transcatheter Aortic Valve Interventions-Registry Investigators
Transcatheter aortic valve implantation: first results from a multi-centre real-world registry
Eur Heart J
 , 
2011
, vol. 
32
 (pg. 
198
-
204
)
30
Moat
NE
Ludman
P
de Belder
MA
Bridgewater
B
Cunningham
AD
Young
CP
Thomas
M
Kovac
J
Spyt
T
MacCarthy
PA
Wendler
O
Hildick-Smith
D
Davies
SW
Trivedi
U
Blackman
DJ
Levy
RD
Brecker
SJ
Baumbach
A
Daniel
T
Gray
H
Mullen
MJ
Long-term outcomes after transcatheter aortic valve implantation in high-risk patients with severe aortic stenosis: the U.K. TAVI (United Kingdom Transcatheter Aortic Valve Implantation) Registry
J Am Coll Cardiol
 , 
2011
, vol. 
58
 (pg. 
2130
-
2138
)
31
Kappetein
AP
Head
SJ
Treede
H
Reichenspurner
H
Mohr
FW
Walther
T
What is the evidence allowing us to state that transcatheter aortic valve replacement via the femoral artery is a more attractive option compared to transapical valve replacement?
EuroIntervention
 , 
2011
, vol. 
7
 (pg. 
903
-
904
)
32
Gilard
M
Eltchaninoff
H
Iung
B
Donzeau-Gouge
P
Chevreul
K
Fajadet
J
Leprince
P
Leguerrier
A
Lievre
M
Prat
A
Teiger
E
Lefevre
T
Himbert
D
Tchetche
D
Carrie
D
Albat
B
Cribier
A
Rioufol
G
Sudre
A
Blanchard
D
Collet
F
Dos Santos
P
Meneveau
N
Tirouvanziam
A
Caussin
C
Guyon
P
Boschat
J
Le Breton
H
Collart
F
Houel
R
Delpine
S
Souteyrand
G
Favereau
X
Ohlmann
P
Doisy
V
Grollier
G
Gommeaux
A
Claudel
JP
Bourlon
F
Bertrand
B
Van Belle
E
Laskar
M
FRANC 2 Investigators
Registry of transcatheter aortic-valve implantation in high-risk patients
N Engl J Med
 , 
2012
, vol. 
366
 (pg. 
1705
-
715
)
33
Makkar
RR
Fontana
GP
Jilaihawi
H
Kapadia
S
Pichard
AD
Douglas
PS
Thourani
VH
Babaliaros
VC
Webb
JG
Herrmann
HC
Bavaria
JE
Kodali
S
Brown
DL
Bowers
B
Dewey
TM
Svensson
LG
Tuzcu
M
Moses
JW
Williams
MR
Siegel
RJ
Akin
JJ
Anderson
WN
Pocock
S
Smith
CR
Leon
MB
PARTNER Trial Investigators
Transcatheter aortic-valve replacement for inoperable severe aortic stenosis
N Engl J Med
 , 
2012
, vol. 
366
 (pg. 
1696
-
1704
)
34
Kodali
SK
Williams
MR
Smith
CR
Svensson
LG
Webb
JG
Makkar
RR
Fontana
GP
Dewey
TM
Thourani
VH
Pichard
AD
Fischbein
M
Szeto
WY
Lim
S
Greason
KL
Teirstein
PS
Malaisrie
SC
Douglas
PS
Hahn
RT
Whisenant
B
Zajarias
A
Wang
D
Akin
JJ
Anderson
WN
Leon
MB
PARTNER Trial Investigators
Two-year outcomes after transcatheter or surgical aortic-valve replacement
N Engl J Med
 , 
2012
, vol. 
366
 (pg. 
1686
-
1695
)
35
Head
SJ
Kaul
S
Bogers
AJ
Kappetein
AP
Non-inferiority study design: lessons to be learned from cardiovascular trials
Eur Heart J
 , 
2012
, vol. 
33
 (pg. 
1318
-
1324
)
36
Reynolds
MR
Magnuson
EA
Wang
K
Lei
Y
Vilain
K
Walczak
J
Kodali
SK
Lasala
JM
O'Neill
WW
Davidson
CJ
Smith
CR
Leon
MB
Cohen
DJ
PARTNER Trial Investigators
Cost-effectiveness of transcatheter aortic valve replacement compared with standard care among inoperable patients with severe aortic stenosis: results from the placement of aortic transcatheter valves (PARTNER) trial (Cohort B)
Circulation
 , 
2012
, vol. 
125
 (pg. 
1102
-
1109
)
37
Van Mieghem
NM
Luthen
C
Oei
F
Schultz
C
Ligthart
J
Kappetein
AP
de Jaegere
PP
Completely percutaneous transcatheter aortic valve implantation through transaxillary route: an evolving concept
EuroIntervention
 , 
2012
, vol. 
7
 (pg. 
1340
-
1342
)
38
Bapat
V
Khawaja
MZ
Attia
R
Narayana
A
Wilson
K
Macgillivray
K
Young
C
Hancock
J
Redwood
S
Thomas
M
Transaortic transcatheter aortic valve implantation using edwards sapien valve: a novel approach
Catheter Cardiovasc Interv
 , 
2012
, vol. 
79
 (pg. 
733
-
740
)
39
Latsios
G
Gerckens
U
Grube
E
Transaortic transcatheter aortic valve implantation: a novel approach for the truly ‘no-access option’ patients
Catheter Cardiovasc Interv
 , 
2010
, vol. 
75
 (pg. 
1129
-
1136
)
40
Modine
T
Lemesle
G
Azzaoui
R
Sudre
A
Aortic valve implantation with the CoreValve ReValving System via left carotid artery access: first case report
J Thorac Cardiovasc Surg
 , 
2010
, vol. 
140
 (pg. 
928
-
929
)
41
Wenaweser
P
Buellesfeld
L
Gerckens
U
Grube
E
Percutaneous aortic valve replacement for severe aortic regurgitation in degenerated bioprosthesis: the first valve in valve procedure using the Corevalve Revalving system
Catheter Cardiovasc Interv
 , 
2007
, vol. 
70
 (pg. 
760
-
764
)
42
Gurvitch
R
Cheung
A
Ye
J
Wood
DA
Willson
AB
Toggweiler
S
Binder
R
Webb
JG
Transcatheter valve-in-valve implantation for failed surgical bioprosthetic valves
J Am Coll Cardiol
 , 
2011
, vol. 
58
 (pg. 
2196
-
2209
)
43
Piazza
N
Bleiziffer
S
Brockmann
G
Hendrick
R
Deutsch
MA
Opitz
A
Mazzitelli
D
Tassani-Prell
P
Schreiber
C
Lange
R
Transcatheter aortic valve implantation for failing surgical aortic bioprosthetic valve: from concept to clinical application and evaluation (part 2)
JACC Cardiovasc Interv
 , 
2011
, vol. 
4
 (pg. 
733
-
742
)
44
Cerillo
AG
Chiaramonti
F
Murzi
M
Bevilacqua
S
Cerone
E
Palmieri
C
Del Sarto
P
Mariani
M
Berti
S
Glauber
M
Transcatheter valve in valve implantation for failed mitral and tricuspid bioprosthesis
Catheter Cardiovasc Interv
 , 
2011
, vol. 
78
 (pg. 
987
-
995
)
45
de Weger
A
Ewe
SH
Delgado
V
Bax
JJ
First-in-man implantation of a trans-catheter aortic valve in a mitral annuloplasty ring: novel treatment modality for failed mitral valve repair
Eur J Cardiothorac Surg
 , 
2011
, vol. 
39
 (pg. 
1054
-
1056
)
46
Gurvitch
R
Cheung
A
Bedogni
F
Webb
JG
Coronary obstruction following transcatheter aortic valve-in-valve implantation for failed surgical bioprostheses
Catheter Cardiovasc Interv
 , 
2011
, vol. 
77
 (pg. 
439
-
444
)
47
Azadani
AN
Jaussaud
N
Matthews
PB
Ge
L
Chuter
TA
Tseng
EE
Transcatheter aortic valves inadequately relieve stenosis in small degenerated bioprostheses
Interact Cardiovasc Thorac Surg
 , 
2010
, vol. 
11
 (pg. 
70
-
77
)
48
Seiffert
M
Conradi
L
Baldus
S
Knap
M
Schirmer
J
Franzen
O
Koschyk
D
Meinertz
T
Reichenspurner
H
Treede
H
Impact of patient-prosthesis mismatch after transcatheter aortic valve-in-valve implantation in degenerated bioprostheses
J Thorac Cardiovasc Surg
 , 
2012
, vol. 
143
 (pg. 
617
-
624
)
49
Soon
JL
Ye
J
Lichtenstein
SV
Wood
D
Webb
JG
Cheung
A
Transapical transcatheter aortic valve implantation in the presence of a mitral prosthesis
J Am Coll Cardiol
 , 
2011
, vol. 
58
 (pg. 
715
-
721
)
50
Bagur
R
Rodes-Cabau
J
Dumont
E
Larochelliere
RD
Doyle
D
Bertrand
OF
Cote
M
Poirier
P
Pibarot
P
Exercise capacity in patients with severe symptomatic aortic stenosis before and six months after transcatheter aortic valve implantation
Am J Cardiol
 , 
2011
, vol. 
108
 (pg. 
258
-
264
)
51
Bagur
R
Rodes-Cabau
J
Dumont
E
De Larochelliere
R
Doyle
D
Pibarot
P
Cote
M
Clavel
MA
Villeneuve
J
Gutierrez
M
Poirier
P
Bertrand
OF
Performance-based functional assessment of patients undergoing transcatheter aortic valve implantation
Am Heart J
 , 
2011
, vol. 
161
 (pg. 
726
-
734
)
52
Gurvitch
R
Wood
DA
Tay
EL
Leipsic
J
Ye
J
Lichtenstein
SV
Thompson
CR
Carere
RG
Wijesinghe
N
Nietlispach
F
Boone
RH
Lauck
S
Cheung
A
Webb
JG
Transcatheter aortic valve implantation: durability of clinical and hemodynamic outcomes beyond 3 years in a large patient cohort
Circulation
 , 
2010
, vol. 
122
 (pg. 
1319
-
1327
)
53
Clavel
MA
Webb
JG
Pibarot
P
Altwegg
L
Dumont
E
Thompson
C
De Larochelliere
R
Doyle
D
Masson
JB
Bergeron
S
Bertrand
OF
Rodes-Cabau
J
Comparison of the hemodynamic performance of percutaneous and surgical bioprostheses for the treatment of severe aortic stenosis
J Am Coll Cardiol
 , 
2009
, vol. 
53
 (pg. 
1883
-
1891
)
54
Head
SJ
Mokhles
MM
Osnabrugge
RL
Pibarot
P
Mack
MJ
Takkenberg
JJ
Bogers
AJ
Kappetein
AP
The impact of prosthesis-patient mismatch on long-term survival after aortic valve replacement: a systematic review and meta-analysis of 34 observational studies comprising 27 186 patients with 133 141 patient-years
Eur Heart J
 , 
2012
, vol. 
33
 (pg. 
1518
-
1529
)
55
O'Brien
SM
Shahian
DM
Filardo
G
Ferraris
VA
Haan
CK
Rich
JB
Normand
SL
DeLong
ER
Shewan
CM
Dokholyan
RS
Peterson
ED
Edwards
FH
Anderson
RP
Society of Thoracic Surgeons Quality Measurement Task Force
The Society of Thoracic Surgeons 2008 cardiac surgery risk models: part 2—isolated valve surgery
Ann Thorac Surg
 , 
2009
, vol. 
88
 (pg. 
S23
-
S42
)
56
Rosenhek
R
Iung
B
Tornos
P
Antunes
MJ
Prendergast
BD
Otto
CM
Kappetein
AP
Stepinska
J
Kaden
JJ
Naber
CK
Acarturk
E
Gohlke–Barwolf
C
ESC working group on valvular heart disease position paper: assessing the risk of interventions in patients with valvular heart disease
Eur Heart J
 , 
2012
, vol. 
33
 (pg. 
822
-
828
)
57
Head
SJ
Bogers
AJ
Serruys
PW
Takkenberg
JJ
Kappetein
AP
A crucial factor in shared decision making: the team approach
Lancet
 , 
2011
, vol. 
377
 pg. 
1836
 
58
Head
SJ
van Mieghem
NM
Serruys
PW
Kappetein
AP
Eeckhout
E
Serruys
PW
Wijns
W
Vahanian
A
Van Sambeek
M
De Palma
R
The Heart Team
Percutaneous Interventional Cardiovascular Medicine
 , 
2012
Toulouse
PCR Publishing
59
Delgado
V
Schuijf
JD
Bax
JJ
Pre-operative aortic valve implantation evaluation: multimodality imaging
EuroIntervention
 , 
2010
, vol. 
6
 (pg. 
G38
-
G47
)
60
Jilaihawi
H
Kashif
M
Fontana
G
Furugen
A
Shiota
T
Friede
G
Makhija
R
Doctor
N
Leon
MB
Makkar
RR
Cross-sectional computed tomographic assessment improves accuracy of aortic annular sizing for transcatheter aortic valve replacement and reduces the incidence of paravalvular aortic regurgitation
J Am Coll Cardiol
 , 
2012
, vol. 
59
 (pg. 
1275
-
1286
)
61
Hamdan
A
Guetta
V
Konen
E
Goitein
O
Segev
A
Raanani
E
Spiegelstein
D
Hay
I
Di Segni
E
Eldar
M
Schwammenthal
E
Deformation dynamics and mechanical properties of the aortic annulus by 4-dimensional computed tomography
J Am Coll Cardiol
 , 
2012
, vol. 
59
 (pg. 
119
-
127
)
62
Gurvitch
R
Webb
JG
Yuan
R
Johnson
M
Hague
C
Willson
AB
Toggweiler
S
Wood
DA
Ye
J
Moss
R
Thompson
CR
Achenbach
S
Min
JK
Labounty
TM
Cury
R
Leipsic
J
Aortic annulus diameter determination by multidetector computed tomography: reproducibility, applicability, and implications for transcatheter aortic valve implantation
JACC Cardiovasc Interv
 , 
2011
, vol. 
4
 (pg. 
1235
-
1245
)
63
Gurvitch
R
Tay
EL
Wijesinghe
N
Ye
J
Nietlispach
F
Wood
DA
Lichtenstein
S
Cheung
A
Webb
JG
Transcatheter aortic valve implantation: lessons from the learning curve of the first 270 high-risk patients
Catheter Cardiovasc Interv
 , 
2011
, vol. 
78
 (pg. 
977
-
984
)
64
Grube
E
Buellesfeld
L
Mueller
R
Sauren
B
Zickmann
B
Nair
D
Beucher
H
Felderhoff
T
Iversen
S
Gerckens
U
Progress and current status of percutaneous aortic valve replacement: results of three device generations of the CoreValve Revalving system
Circ Cardiovasc Interv
 , 
2008
, vol. 
1
 (pg. 
167
-
175
)
65
Alli
OO
Booker
JD
Lennon
RJ
Greason
KL
Rihal
CS
Holmes
DR
Jr
Transcatheter aortic valve implantation: assessing the learning curve
JACC Cardiovasc Interv
 , 
2012
, vol. 
5
 (pg. 
72
-
79
)
66
Nuis
RJ
van Mieghem
NM
van der Boon
RM
van Geuns
RJ
Schultz
CJ
Oei
FB
Galema
TW
Raap
GB
Koudstaal
PJ
Geleijnse
ML
Kappetein
AP
Serruys
PW
de Jaegere
PP
Effect of experience on results of transcatheter aortic valve implantation using a Medtronic CoreValve System
Am J Cardiol
 , 
2011
, vol. 
107
 (pg. 
1824
-
1829
)