Abstract

Purpose

To develop a method of coronary flow reserve (CFR) calculation derived from three-dimensional (3D) coronary angiographic parameters and intracoronary pressure data during fractional flow reserve (FFR) measurement.

Methods

Altogether 19 coronary arteries of 16 native and 3 stented vessels were reconstructed in 3D. The measured distal intracoronary pressures were corrected to the hydrostatic pressure based on the height differences between the levels of the vessel orifice and the sensor position. Classical fluid dynamic equations were applied to calculate the flow during the resting state and vasodilatation on the basis of morphological data and intracoronary pressure values. 3D-derived coronary flow reserve (CFRp-3D) was defined as the ratio between the calculated hyperemic and the resting flow and was compared to the CFR values simultaneously measured by the Doppler sensor (CFRDoppler).

Results

Haemodynamic calculations using the distal coronary pressures corrected for hydrostatic pressures showed a strong correlation between the individual CFRp-3D values and the CFRDoppler measurements (r=0.89, p<0.0001). Hydrostatic pressure correction increased the specificity of the method from 46.1% to 92.3% for predicting an abnormal CFRDoppler <2.

Conclusions

CFRp-3D calculation with hydrostatic pressure correction during FFR measurement facilitates a comprehensive haemodynamic assessment, supporting the complex evaluation of macro- and microvascular coronary artery disease.

Funding Acknowledgement

Type of funding sources: Public Institution(s). Main funding source(s): University of Debrecen

Calculation of CFR and MRR

Calculation of CFR and MRR

This content is only available as a PDF.
This article is published and distributed under the terms of the Oxford University Press, Standard Journals Publication Model (https://academic.oup.com/journals/pages/open_access/funder_policies/chorus/standard_publication_model)