Aims

In the setting of paroxysmal atrial fibrillation (AF), there are no available data comparing the mid-term outcome of patients undergoing pulmonary vein isolation (PVI) catheter ablation using contact-force (CF)-guided radiofrequency (RF) vs. second-generation balloon cryotherapy.

Methods and results

Prospective single-centre evaluation, carried out from March 2011 to February 2013, comparing CF radiofrequency (Thermocool® SmartTouch™, Biosense Webster, Inc.) (CF group) with cryoballoon ablation (Arctic Front Advance™ 28 mm cryoballoon, Medtronic, Inc.) (CB group), in regards to procedural safety and efficacy, as well as recurrence at 12 months. Overall, 150 consecutive patients were enrolled (75 in each group). The characteristics of patients of both the groups were similar (61.2 ± 9.9 years, women 25.3%, mean AF duration 4.1 ± 4.0 years, mean CHA2DS2-VASc score 1.4 ± 1.3, mean HAS-BLED 1.4 ± 0.6). Duration of the procedure was significantly lower in the CF group (110.7 ± 32.5 vs. 134.5 ± 48.3 min, P = 0.001), with a lower duration of fluoroscopy (21.5 ± 8.5 vs. 25.3 ± 9.9 min, P = 0.017) and X-ray exposure (4748 ± 2411 cGy cm² vs. 7734 ± 5361 cGy cm², P = 0.001). In contrast, no significant difference was found regarding significant procedural complication (2.7 vs. 1.3% in CF and CB groups, respectively; P = 0.56), and PVI was eventually achieved in all cases. At 12 months, AF recurrence occurred in 11 patients (14.7%) in the CB group and in 9 patients (12.0%) in the CF group (HR = 1.20 95% CI 0.50–2.90; log rank P = 0.682).

Conclusions

Our preliminary findings suggest that CF-guided radiofrequency and cryotherapy present very similar performances in the setting of paroxysmal AF catheter ablation.

What's new?

  • Mid-term outcome of patients undergoing paroxysmal atrial fibrillation (AF) ablation with contact-force-guided radiofrequency

  • Mid-term outcome of patients undergoing paroxysmal AF ablation with second-generation cryoballoon.

  • Prospective comparison of the safety and efficacy of these two novel technologies.

Introduction

Catheter ablation has been shown to be effective in treating patients with symptomatic paroxysmal atrial fibrillation (AF).1 Pulmonary vein isolation (PVI) remains the corner stone of the strategy.2 However, AF recurrences after a single procedure remain relatively frequent and in most of cases are related to PV reconnections.3 This may reflect the lack of effectiveness in achieving transmural, continuous and long-lasting lesions when performing PVI.4,5

The contact force (CF) between catheter tip and target tissue has been shown to be a major influencing factor in providing effective tissue lesion.610 Accordingly, novel technologies for AF catheter ablation have recently focused on the optimization of the contact between both interfaces. On the one hand, two new radiofrequency catheters including the CF-sensing technology have been recently developed and commercialized allowing continuous CF monitoring during ablation.11 On the other hand, second generation of cryoballoon (CB), the Arctic Front Advance™, has been developed to optimize lesions in various settings of PV anatomies.1214 The extent to which the CF technology may be potentially of additional value compared with cryotherapy (in terms of mid-term outcome) has not been evaluated so far.

In the present analysis, we prospectively compared the safety and efficacy (procedural and 12 months) of CF technology with second-generation CB in the setting of paroxysmal AF catheter ablation.

Methods

Setting and patient population

In this single-centre (Clinique Pasteur, France), non-randomized, prospective study, we compared procedural and 12-month outcomes of patients who underwent PVI in the setting of paroxysmal AF according to two different technologies: (i) radiofrequency ablation with the Thermocool® SmartTouch™ (Biosense Webster, Inc.) catheter (CF group), from March 2011 to June 2012; (ii) Arctic Front Advance™ 28 mm CB (Medtronic, Inc.) (CB group) from July 2012 to February 2013. Eligible patients were patients with paroxysmal AF refractory to at least one anti-arrhythmic drug from Class I or III undergoing a first PVI procedure. Exclusion criteria were in need of additional lines (i.e. roof, mitral isthmus) or ablation of complex-fractionated electrograms during the procedure. Of note, in our centre, the operators are dedicated exclusively to one of the techniques (two different operators for both CF and CB techniques) with a relatively high activity (>100 AF ablation procedures/year/operator). Patient selection for one or the other group only depends on the electrophysiology physician they consulted before being referred for the procedure.

All patients provided an informed consent prior to the procedure. The study complied with the Declaration of Helsinki and the research protocol was approved by the local ethics committee.

Procedural details on ablation procedure

In all cases, computed tomographic scanning of the left atrium was performed 24 h prior to the procedure to assess PV anatomy and exclude the presence of left atrial thrombus. The procedure was performed under general anaesthesia like every AF ablation procedure in our centre. Using a transfemoral venous approach, a quadripolar catheter was placed in the coronary sinus. A single transseptal puncture was performed under fluoroscopic guidance. Transesophageal echocardiographic guidance was used only in case of failure of the usual fluoroscopic-guided approach. Upon completion of the transseptal puncture, patients received intravenous heparin to maintain an activated clotting time of 300–350 s. All patients underwent AF ablation were treated with interrupted vitamin-K antagonists and a peri-procedural bridging with subcutaneous heparin. Further details on the anticoagulation protocol have already been described.15 No oesophageal monitoring was done during the procedure.

Contact-force group

A circumferential mapping catheter (Lasso™; Biosense Webster©) was introduced into the left atrium using an 8.5 French long sheath. The Lasso catheter was used to collect left atrial geometry using the three-dimensional electroanatomic mapping system Carto 3®. After completion of left atrial geometry, the Lasso catheter was removed and an ablation catheter was introduced into the left atrium through the same transseptal sheath. Thermocool® SmartTouch™ (Biosense Webster©) was introduced through the same transseptal sheath. Initial circumferential PV ablation was performed by systematic radiofrequency application around the PV ostia, consisting of an encirclement of ipsilateral pairs of PV antra (no >2 cm from the ostium on the posterior wall on both sides, and anterior aspect of right PVs, and guided by the ridge between the left PVs and the appendage) without additional adjunctive left atrial ablation. Power was limited to 30 W at anterior, superior, and inferior sites (flow rate, 17–20 mL/min) and 25 W at all posterior sites (flow rate, 17 mL/min), with temperature limited to 48°C for each lesion. Power was not adjusted according to CF. The novel CF-sensing catheter and viewing platform was used (Thermocool® SmartTouch™, Biosense Webster©) to continuously assess catheter–tissue contact. Contact-force data were available to the operator throughout the procedure. The aim was to achieve a CF of at least 10 grams (mean) with a vector perpendicular to the tissue. The upper limit defined was 50 g force. These values were chosen based on animal studies, shown to be effective in lesion formation while avoiding perforation.7,9 Pulmonary vein isolation was first performed ‘anatomically’ (i.e. without the Lasso catheter, using an exclusive anatomic approach), the Lasso catheter being used only after completion of anatomic PVI to confirm full PV disconnection. This was done by testing both entrance and exit block (bidirectional block), with a waiting period of 20 min after the last radiofrequency application. In the event, complete disconnection was not achieved (failure of either exit or entrance block), the ablation catheter was reintroduced into the left atrium with the Lasso remaining in place using a double wire technique through the transseptal sheath. Thereafter, PVI was completed with Lasso guidance to eliminate all points of residual PV connection.

Cryoballoon group

A 14 French deflectable sheath (FlexCath® Medtronic©) was introduced into the left atrium after a single transseptal puncture. Then Arctic Front Advance™ balloon was introduced in the sheath, inflated, and advanced to the ostium of each PV and ablation of pulmonary vein antra was performed with two applications of 240 s per vein. Occlusion of each vein was assessed either with venous angiography or with venous pressure curve at the discretion of the operator.16 Continuous monitoring of the phrenic nerve during ablation of the PVs was systematically performed by pacing the right phrenic nerve with a quadripolar catheter in the superior vena cava.

Pulmonary vein isolation was either assessed continuously using the circular Achieve® Catheter (Medtronic©) during CB freezing or traditionally using a circular catheter (Lasso®, Biosense Webster©) after two applications. If the PV remained connected, additional applications were performed using different angulations. Pulmonary vein isolation was finally checked 20 min after the last CB ablation.

Standardized follow-up

After the index procedure, patients were followed for a total of 12 months. Patients were evaluated before hospital discharge, as well as at 1, 3, 6, 9, and 12 months after the procedure. A systematic transthoracic echocardiography and 24-h Holter monitoring were performed before discharge. Information collected during follow-up included a 12-lead electrocardiogram (ECG) and a 24-h Holter monitoring at each visit. No antiarrhythmic medication was prescribed after catheter ablation, and the first 3 months post-procedure was considered as blanking period.2 If there was documented recurrence of symptomatic AF during this time interval and the patient required antiarrhythmic drug therapy, a previously ineffective but tolerated Class I or Class III (sotalol) drug was the preferred option. Anticoagulation strategy after the first 3 months was based on the CHA2DS2Vasc and HAS-BLED scores.2

Procedural and 12-month endpoints

Procedural endpoint was PV isolation confirmed by entry and exit block after a waiting time of 20 min. The specific differences of the two used therapeutic approaches are described above. The following procedural safety and efficacy endpoints were assessed, using the criteria proposed by Sorgente et al.,17 especially the primary procedural efficacy endpoint proportion of effective PVI. Regarding the mid-term follow-up, the endpoint was defined by the rate of AF recurrence, defined as any symptomatic or asymptomatic atrial arrhythmia lasting >30 s after the blanking period during the year after catheter ablation. This was actively ascertained by Holter monitoring at 1, 3, 6, 9, and 12 months or by 12-lead ECG in the case of symptomatic palpitation at clinical interview, in the absence of antiarrhythmic therapy. Patients with recurrence of AF during the blanking period with no response to cardioversion (either pharmacological or direct-current) were classified as having a relapse (only one case in the CB group).

Statistical analysis

Comparisons were performed between the two treatment groups. χ2 was used for nominal variables and Student's t-test was used for comparison of continuous variables, where appropriate; Levene's test was used to check the homogeneity of variance; equivalent non-parametric tests were used when Kolmogorov–Smirnov was in favour of absence of normal distribution. Results with P < 0.05 were regarded as significant. Variables that differed at baseline between the two treatment groups and their impact on sinus rhythm maintenance were assessed using Cox regression (using the forward stepwise method likelihood ratio; probability for stepwise = 0.05). Cox regression was also performed for assessing the predictors of relapse. Kaplan–Meier curves were traced for comparing sinus rhythm maintenance among the two treatment strategies and the log-rank test was used for assessing existing differences. Our hypothesis was that no differences would be observed among the two therapeutic approaches as regards AF relapse. An effect size of 0.2 was estimated using contingency tables assuming two equal samples and a 25% relapse rate, equal in H0 and with a 50% increase in one of the treatment groups as the alternative hypothesis. Thus, the computed achieved power (β) of the sample assuming the effect size of 0.2, nominal α of 0.05, and a sample size of 150 was estimated to be 0.7. PASW Statistics version 18.0 was used for descriptive and inferential statistical analysis.

Results

Patient characteristics

Population characteristics are detailed in Table 1. The mean age of patients was 61.2 ± 9.9 years. The proportion of women was 25.3%; mean AF duration was 4.1 ± 4.0 years. Mean CHADS2 score was 0.8 ± 0.9, CHA2DS2-VASc score 1.4 ± 1.3, and HAS-BLED 1.4 ± 0.6 without significant differences between the two groups. Overall, no significant baseline differences were observed between the CF and CB groups, except for body mass index (28.3 ± 4.6 for CB group vs. 26.5 ± 3.9 for CF group, P = 0.011) and creatinine clearance assessed through the modification of diet in renal disease (81.8 ± 22.5 mL/min in the CB group vs. 73.2 ± 14.9 mL/min in the RF group, P = 0.001).

Table 1

Baseline characteristics of patients

Overall (n = 150)CF group (n = 75)CB group (n = 75)P
Age, years61.2 ± 9.962.5 ± 8.959.9 ± 10.60.111
Female gender25.3% (38)24.0% (18)26.7% (20)0.707
BMI27.4 ± 4.326.5 ± 3.928.2 ± 4.60.015
AF duration, years4.1 ± 4.04.4 ± 4.13.8 ± 4.00.321
HF4.7% (7)2.7% (2)6.7% (5)0.246
HTN41.3% (62)48.0% (36)34.7% (26)0.097
DM6.0% (9)4.0% (3)8.0% (6)0.302
Stroke/TIA7.3% (11)10.7% (8)4.0% (3)0.117
Vascular disease10.0% (15)9.3% (7)10.7% (8)0.785
CHADS20.8 ± 0.90.9 ± 0.90.7 ± 0.90.162
CHA2DS2-VASc1.4 ± 1.31.5 ± 1.31.3 ± 1.30.308
HAS-BLED1.4 ± 0.61.5 ± 0.61.4 ± 0.50.243
Sleep apnoea8.7% (13)5.3% (4)12.0% (9)0.147
Indexed LAV, cm341.4 ± 13.439.5 ± 11.342.8 ± 15.20.127
LVEF, %65.0 ± 6.565.5 ± 5.664.4 ± 7.40.323
Haemoglobin, g/dL14.8 ± 1.314.8 ± 1.214.8 ± 1.40.954
Clearance MDRD, mL/min77.5 ± 19.573.2 ± 14.981.8 ± 22.50.006
CRP, mg/L5.0 ± 14.85.8 ± 19.84.2 ± 6.60.511
Overall (n = 150)CF group (n = 75)CB group (n = 75)P
Age, years61.2 ± 9.962.5 ± 8.959.9 ± 10.60.111
Female gender25.3% (38)24.0% (18)26.7% (20)0.707
BMI27.4 ± 4.326.5 ± 3.928.2 ± 4.60.015
AF duration, years4.1 ± 4.04.4 ± 4.13.8 ± 4.00.321
HF4.7% (7)2.7% (2)6.7% (5)0.246
HTN41.3% (62)48.0% (36)34.7% (26)0.097
DM6.0% (9)4.0% (3)8.0% (6)0.302
Stroke/TIA7.3% (11)10.7% (8)4.0% (3)0.117
Vascular disease10.0% (15)9.3% (7)10.7% (8)0.785
CHADS20.8 ± 0.90.9 ± 0.90.7 ± 0.90.162
CHA2DS2-VASc1.4 ± 1.31.5 ± 1.31.3 ± 1.30.308
HAS-BLED1.4 ± 0.61.5 ± 0.61.4 ± 0.50.243
Sleep apnoea8.7% (13)5.3% (4)12.0% (9)0.147
Indexed LAV, cm341.4 ± 13.439.5 ± 11.342.8 ± 15.20.127
LVEF, %65.0 ± 6.565.5 ± 5.664.4 ± 7.40.323
Haemoglobin, g/dL14.8 ± 1.314.8 ± 1.214.8 ± 1.40.954
Clearance MDRD, mL/min77.5 ± 19.573.2 ± 14.981.8 ± 22.50.006
CRP, mg/L5.0 ± 14.85.8 ± 19.84.2 ± 6.60.511

BMI, body mass index; HF, heart failure; HTN, hypertension; DM, diabetes mellitus; TIA, transient ischaemic attack; LAV, left atrial volume; LVEF, left ventricle ejection fraction; CRP, C-reactive protein.

Table 1

Baseline characteristics of patients

Overall (n = 150)CF group (n = 75)CB group (n = 75)P
Age, years61.2 ± 9.962.5 ± 8.959.9 ± 10.60.111
Female gender25.3% (38)24.0% (18)26.7% (20)0.707
BMI27.4 ± 4.326.5 ± 3.928.2 ± 4.60.015
AF duration, years4.1 ± 4.04.4 ± 4.13.8 ± 4.00.321
HF4.7% (7)2.7% (2)6.7% (5)0.246
HTN41.3% (62)48.0% (36)34.7% (26)0.097
DM6.0% (9)4.0% (3)8.0% (6)0.302
Stroke/TIA7.3% (11)10.7% (8)4.0% (3)0.117
Vascular disease10.0% (15)9.3% (7)10.7% (8)0.785
CHADS20.8 ± 0.90.9 ± 0.90.7 ± 0.90.162
CHA2DS2-VASc1.4 ± 1.31.5 ± 1.31.3 ± 1.30.308
HAS-BLED1.4 ± 0.61.5 ± 0.61.4 ± 0.50.243
Sleep apnoea8.7% (13)5.3% (4)12.0% (9)0.147
Indexed LAV, cm341.4 ± 13.439.5 ± 11.342.8 ± 15.20.127
LVEF, %65.0 ± 6.565.5 ± 5.664.4 ± 7.40.323
Haemoglobin, g/dL14.8 ± 1.314.8 ± 1.214.8 ± 1.40.954
Clearance MDRD, mL/min77.5 ± 19.573.2 ± 14.981.8 ± 22.50.006
CRP, mg/L5.0 ± 14.85.8 ± 19.84.2 ± 6.60.511
Overall (n = 150)CF group (n = 75)CB group (n = 75)P
Age, years61.2 ± 9.962.5 ± 8.959.9 ± 10.60.111
Female gender25.3% (38)24.0% (18)26.7% (20)0.707
BMI27.4 ± 4.326.5 ± 3.928.2 ± 4.60.015
AF duration, years4.1 ± 4.04.4 ± 4.13.8 ± 4.00.321
HF4.7% (7)2.7% (2)6.7% (5)0.246
HTN41.3% (62)48.0% (36)34.7% (26)0.097
DM6.0% (9)4.0% (3)8.0% (6)0.302
Stroke/TIA7.3% (11)10.7% (8)4.0% (3)0.117
Vascular disease10.0% (15)9.3% (7)10.7% (8)0.785
CHADS20.8 ± 0.90.9 ± 0.90.7 ± 0.90.162
CHA2DS2-VASc1.4 ± 1.31.5 ± 1.31.3 ± 1.30.308
HAS-BLED1.4 ± 0.61.5 ± 0.61.4 ± 0.50.243
Sleep apnoea8.7% (13)5.3% (4)12.0% (9)0.147
Indexed LAV, cm341.4 ± 13.439.5 ± 11.342.8 ± 15.20.127
LVEF, %65.0 ± 6.565.5 ± 5.664.4 ± 7.40.323
Haemoglobin, g/dL14.8 ± 1.314.8 ± 1.214.8 ± 1.40.954
Clearance MDRD, mL/min77.5 ± 19.573.2 ± 14.981.8 ± 22.50.006
CRP, mg/L5.0 ± 14.85.8 ± 19.84.2 ± 6.60.511

BMI, body mass index; HF, heart failure; HTN, hypertension; DM, diabetes mellitus; TIA, transient ischaemic attack; LAV, left atrial volume; LVEF, left ventricle ejection fraction; CRP, C-reactive protein.

Procedural results

Procedural data are detailed in Table 2. Duration of the procedure was significantly lower in the CF group with 110.7 ± 32.5 min vs. 134.5 ± 48.3 min in the CB group (P = 0.001) (Table 2). Consequently, duration of fluoroscopy and X-ray dosage were also significantly lower in the CF group (21.5 ± 8.5 vs. 25.3 ± 9.9 min; P = 0.017 and 4748 ± 2411 cGy cm² vs. 7734 ± 5361 cGy cm², P = 0.001) (Figures 1 and 2).
Table 2

Procedural data

Overall (n = 150)CF group (n = 75)CB group (n = 75)P
Presence of left commun trunk, %18% (27)24 (15%)16 (12%)0.505
Duration of procedure, min122.0 ± 42.3110.7 ± 32.5134.5 ± 48.30.001
Duration of fluoroscopy, min23.4 ± 9.421.5 ± 8.525.3 ± 9.90.017
X-Ray dosage, cGy cm26123 ± 42964748 ± 24117734 ± 53610.001
Procedural data—cryo patients N = 75Procedural data—CF patients N = 75
Utilization of the achieve54.7% (41)RF duration, s1954 ± 805
Total number of applications8.5 ± 1.0Average impedance, Ohm145 ± 14
Time of cryoenergy, s1933 ± 175Average power, W22.9 ± 2.1
Average minimum temperature, °C−52.9 ± 5.0Average CF, g19.7 ± 38
Overall (n = 150)CF group (n = 75)CB group (n = 75)P
Presence of left commun trunk, %18% (27)24 (15%)16 (12%)0.505
Duration of procedure, min122.0 ± 42.3110.7 ± 32.5134.5 ± 48.30.001
Duration of fluoroscopy, min23.4 ± 9.421.5 ± 8.525.3 ± 9.90.017
X-Ray dosage, cGy cm26123 ± 42964748 ± 24117734 ± 53610.001
Procedural data—cryo patients N = 75Procedural data—CF patients N = 75
Utilization of the achieve54.7% (41)RF duration, s1954 ± 805
Total number of applications8.5 ± 1.0Average impedance, Ohm145 ± 14
Time of cryoenergy, s1933 ± 175Average power, W22.9 ± 2.1
Average minimum temperature, °C−52.9 ± 5.0Average CF, g19.7 ± 38
Table 2

Procedural data

Overall (n = 150)CF group (n = 75)CB group (n = 75)P
Presence of left commun trunk, %18% (27)24 (15%)16 (12%)0.505
Duration of procedure, min122.0 ± 42.3110.7 ± 32.5134.5 ± 48.30.001
Duration of fluoroscopy, min23.4 ± 9.421.5 ± 8.525.3 ± 9.90.017
X-Ray dosage, cGy cm26123 ± 42964748 ± 24117734 ± 53610.001
Procedural data—cryo patients N = 75Procedural data—CF patients N = 75
Utilization of the achieve54.7% (41)RF duration, s1954 ± 805
Total number of applications8.5 ± 1.0Average impedance, Ohm145 ± 14
Time of cryoenergy, s1933 ± 175Average power, W22.9 ± 2.1
Average minimum temperature, °C−52.9 ± 5.0Average CF, g19.7 ± 38
Overall (n = 150)CF group (n = 75)CB group (n = 75)P
Presence of left commun trunk, %18% (27)24 (15%)16 (12%)0.505
Duration of procedure, min122.0 ± 42.3110.7 ± 32.5134.5 ± 48.30.001
Duration of fluoroscopy, min23.4 ± 9.421.5 ± 8.525.3 ± 9.90.017
X-Ray dosage, cGy cm26123 ± 42964748 ± 24117734 ± 53610.001
Procedural data—cryo patients N = 75Procedural data—CF patients N = 75
Utilization of the achieve54.7% (41)RF duration, s1954 ± 805
Total number of applications8.5 ± 1.0Average impedance, Ohm145 ± 14
Time of cryoenergy, s1933 ± 175Average power, W22.9 ± 2.1
Average minimum temperature, °C−52.9 ± 5.0Average CF, g19.7 ± 38
Catheter ablation using the CF technology (Thermocool® SmartTouch™, Biosense Webster, Inc.). The objective was to have the ablation catheter oriented the most perpendicular to the wall with a force of at least 10 grams. Here the posterior part of the left PVI.
Figure 1

Catheter ablation using the CF technology (Thermocool® SmartTouch™, Biosense Webster, Inc.). The objective was to have the ablation catheter oriented the most perpendicular to the wall with a force of at least 10 grams. Here the posterior part of the left PVI.

Twelve-month endpoint and procedural data.
Figure 2

Twelve-month endpoint and procedural data.

In both groups, all PVs were disconnected at the end of the procedure. No complementary RF applications were needed to achieve complete PVI in the CB group. Mean CF observed in the CF group was 19.7 g+/−38.

No peri-procedural deaths or thromboembolic events were observed. Major bleeding occurred in two patients in the CF group (groin haematoma and upper gastrointestinal bleeding with the need of red-blood cell transfusion) and in one patient in the CB group (pseudaneurysm requiring fibrin injection) (Table 3). Transient phrenic nerve palsy occurred in 13 (17.3%) patients in the CB group, with normal diaphragmatic function at the end of the procedure.

Table 3

Procedural outcome

Overall (n = 150)CF group (n = 75)CB group (n = 75)P
Major bleeding2.0% (3)2.7% (2)1.3% (1)0.560
Thromboembolic events0% (0)0% (0)0% (0)NA
Peri-procedural deaths0% (0)0% (0)0% (0)NA
Relapse post-blanking13.3% (20)12.0% (9)14.7% (11)0.608
Overall (n = 150)CF group (n = 75)CB group (n = 75)P
Major bleeding2.0% (3)2.7% (2)1.3% (1)0.560
Thromboembolic events0% (0)0% (0)0% (0)NA
Peri-procedural deaths0% (0)0% (0)0% (0)NA
Relapse post-blanking13.3% (20)12.0% (9)14.7% (11)0.608
Table 3

Procedural outcome

Overall (n = 150)CF group (n = 75)CB group (n = 75)P
Major bleeding2.0% (3)2.7% (2)1.3% (1)0.560
Thromboembolic events0% (0)0% (0)0% (0)NA
Peri-procedural deaths0% (0)0% (0)0% (0)NA
Relapse post-blanking13.3% (20)12.0% (9)14.7% (11)0.608
Overall (n = 150)CF group (n = 75)CB group (n = 75)P
Major bleeding2.0% (3)2.7% (2)1.3% (1)0.560
Thromboembolic events0% (0)0% (0)0% (0)NA
Peri-procedural deaths0% (0)0% (0)0% (0)NA
Relapse post-blanking13.3% (20)12.0% (9)14.7% (11)0.608

Twelve-month follow-up

At 12-month follow-up, AF recurrence occurred in 11 patients (14.7%) in the CB group and in 9 patients (12.0%) in the CF group (Univariate Cox Regression HR = 1.20 95% CI 0.50–2.90; P = 0.681; from Kaplan–Meier curves, log rank P = 0.988) (Figure 3). All recurrences were symptomatic and confirmed by a ≥30 s ECG or Holter tracing. A redo procedure was performed until March 2014 in eight patients in the CF group, and in two in the CB group. Although two patients in the CF group did not present any PV reconnection, the average number of reconnected veins was similar in patients treated with CF or CB: 1.6 ± 1.3 vs. 1.5 ± 0.7; P = 0.902, respectively.
Kaplan–Meier survival curve—proportion of patients free of AF during the 12-month follow-up (3-month blanking period).
Figure 3

Kaplan–Meier survival curve—proportion of patients free of AF during the 12-month follow-up (3-month blanking period).

On Cox regression, assessing all variables that on univariate analysis presented P < 0.1 regarding the existence of AF relapse (AF duration in years, heart failure, hypertension, previous stroke or transient ischaemic attack, CHADS2, CHA2DS2-VASc, and HAS-BLED), only AF duration (HR per year = 1.14; CI 95% 1.03–1.26; P = 0.014), heart failure (HR = 6.42; CI 95% 2.05–20.16; P = 0.001) and previous stroke or transient ischaemic attack (HR = 5.85; CI 95% 1.81–18.92; P = 0.003) were independent predictors of AF relapse (see Supplementary material online).

Discussion

Our data suggest that paroxysmal AF patients who underwent ablation using either the novel Artic Front Advance™ CB or the SmartTouch™ CF catheter present similar propensity of remaining free from AF recurrence at 12 months out of any anti-arrhythmic drugs. At the best of our knowledge, this is the first evaluation of these two newly available technologies for AF catheter ablation.

Pulmonary vein isolation is known to be particularly challenging due to the relative difficulty in maintaining a good contact with the tissue all over the encirclement, leading CF technology particularly welcome in this setting. Also, achieving optimal CF since the first application is of particular importance, in order to limit tissue oedema formation. Recent studies have reported that using a CF-sensing catheter leads to overall success and reduces procedural time, and X-ray exposure.1820 As we began AF ablation with the CF-sensing catheter in March 2011, very few studies determining a cut-off value had been published, except in the animals. We determined this minimal value of 10 g with the goal to be safe and effective, according to our preliminary experience. The recently published EFFICAS I study has suggested that a minimum CF of 20 g per application appears to be an interesting cut-off to improve mid-term clinical outcomes,21 that' is this target value was enhanced to 20 g, in line with the average CF measured in our patients (19.7 g).22 Regarding CB ablation, it has recently been reported that the new generation CB improves the efficiency of the procedure reducing the time to PV isolation, procedural time, and overall success compared with the first-generation balloons.2325 Our findings provide additional data showing that CB ablation in paroxysmal AF ablation is a relatively safe and effective technique. Phrenic nerve palsy has been described as the most frequent complication, although the vast majority of them revert shortly. We observed longer procedural and fluoroscopy times in the CB group compared with the CF group; of note, we performed two systematic 240 s applications on each PV in the CB group, whereas recent studies suggest that application time could be probably reduced without interacting with the efficacy of PVI.2628 The short time with CF should be interpreted with some caution. Because our centre has got a strong expertise in the technique, and the time with radiofrequency use is more likely to be influenced by level of operator's experience (in contrast with a more ‘standardised’ time using CB), differences in times between the two techniques could be less significant in less experienced centres. Indeed, CB ablation is usually shorter and more reproducible than radiofrequency ablation in the setting of paroxysmal AF ablation.29,30

Although preliminary experiences comparing cryotherapy with radiofrequency (non-CF) ablation (in terms of acute procedural outcome) are available, to the best of our knowledge, we provide here the first evaluation using mid-term outcomes. The recent results from the German Ablation Registry have suggested that (first-generation) CB ablation and (non-CF) radiofrequency ablation presented a similar efficacy in terms of acute success, in the setting of paroxysmal AF ablation.31 Another recent study has been carried out in a large population of 396 patients from 2008 to 2011.29 The results have shown that the two techniques presented similar peri-procedural efficacy and safety. The investigators have observed a lower proportion of phrenic nerve injury than more recently found with the Arctic Front Advance™.32 In contrast with our results, the investigators of the randomized COR trial have found that CB ablation was inferior to irrigated radiofrequency (non-CF) catheters.33 However, of note, in our study, all PVs were disconnected at the end of the procedure, especially in the CB group, whereas in the COR trial, only 83% of PVs remained disconnected in the CB group. This may be potentially the result of the two systematic 300 s applications, without any additional application in the case of residual veno-atrial conduction. Secondly, in the COR trial, only the first-generation CB was used, which, as previously mentioned, seems to be less effective.

We acknowledge several limitations. First, this is not a randomized trial and inclusion of the patients for one or the other group is subject to bias, despite the similarity of baseline variables that was observed. Secondly, regarding the CB group, PV occlusion verification was checked differently: one operator used pressure-guide while the other used contrast injection. Thirdly, this is a single centre study, from a centre performing more than 500 AF ablation procedures per year. The extent to which these results may apply to less experienced centres needs further investigation. Fourthly, systematic monitoring using an implantable loop recorder might have documented higher rate of asymptomatic recurrence. However, the relative proportion of patients with exclusively asymptomatic AF is not expected to be systematically different between the two groups, and finally the likelihood that this bias was differentially low.

Conclusion

Our findings suggest that CF real-time assessment using a SmartTouch™ catheter and CB ablation using the novel Artic Front Advance™ display a very similar procedural efficacy and safety. More importantly, our results also suggest that the mid-term effectiveness profile as regards the 12-month recurrence rate (almost 85% of patients remaining free of AF without anti-arrhythmic drugs) is highly similar between both the groups. Further randomized and multicentric evaluations are needed to confirm these preliminary results in order to finally identify specific subgroups more likely to benefit from one or the other technique.

Supplementary material

Supplementary material is available at Europace online.

Conflict of interest: S.B. has received fees as a consultant for Medtronic and Boston Scientific. J.P. Albenque is a consultant for Biosense Webster and St. Jude Medical. Other authors: No conflict of interest.

References

1

Raviele
A
Natale
A
Calkins
H
Camm
JA
Cappato
R
Ann Chen
S
et al.
Venice chart international consensus document on atrial fibrillation ablation: 2011 update. Venice Chart
J Cardiovasc Electrophysiol
2012
23
890
923

2

Camm
AJ
Lip
GY
De Caterina
R
Savelieva
I
Atar
D
Hohnloser
SH
et al.
2012 focused update of the ESC Guidelines for the management of atrial fibrillation: an update of the 2010 ESC Guidelines for the management of atrial fibrillation—developed with the special contribution of the European Heart Rhythm Association
Europace
2012
14
1385
413

3

Ouyang
F
Antz
M
Ernst
S
Hachiya
H
Mavrakis
H
Deger
FT
et al.
Recovered pulmonary vein conduction as a dominant factor for recurrent atrial tachyarrhythmias after complete circular isolation of the pulmonary veins: lessons from double Lasso technique
Circulation
2005
111
127
35

4

Verma
A
Kilicaslan
F
Pisano
E
Marrouche
NF
Fanelli
R
Brachmann
J
et al.
Response of atrial fibrillation to pulmonary vein antrum isolation is directly related to resumption and delay of pulmonary vein conduction
Circulation
2005
112
627
35

5

Ouyang
F
Tilz
R
Chun
J
Schmidt
B
Wissner
E
Zerm
T
et al.
Long-term results of catheter ablation in paroxysmal atrial fibrillation: lessons from a 5-year follow-up
Circulation
2010
122
2368
77

6

Haines
DE
Determinants of lesion size during radiofrequency catheter ablation: the role of electrode-tissue contact pressure and duration of energy delivery
J Cardiovasc Electrophysiol
1991
2
509
15

7

Yokoyama
K
Nakagawa
H
Shah
DC
Lambert
H
Leo
G
Aeby
N
et al.
Novel contact force sensor incorporated in irrigated radiofrequency ablation catheter predicts lesion size and incidence of steam pop and thrombus
Circ Arrhythm Electrophysiol
2008
1
354
62

8

Shah
DC
Lambert
H
Nakagawa
H
Langenkamp
A
Aeby
N
Leo
G
Area under the real-time contact force curve (force-time integral) predicts radiofrequency lesion size in an in vitro contractile model simulating beating heart
J Cardiovasc Electrophysiol
2010
21
1038
43

9

Thiagalingam
A
D'Avila
A
Foley
L
Luis Guerrero
J
Lambert
H
Leo
G
et al.
Importance of catheter contact force during irrigated radiofrequency ablation: evaluation in a porcine ex vivo model using a force-sensing catheter
J Cardiovasc Electrophysiol
2010
21
806
11

10

Squara
F
Latcu
DG
Massaad
Y
Mahjoub
M
Bun
SS
Saoudi
N
Contact force and force-time integral in atrial radiofrequency ablation predict transmurality of lesions
Europace
2014
16
660
7

11

Kuck
KH
Reddy
VY
Schmidt
B
Natale
A
Neuzil
P
Saoudi
N
et al.
A novel radiofrequency ablation catheter using contact force sensing: Toccata study
Heart Rhythm
2012
9
18
23

12

Liu
CF
Pulmonary vein reconnection after cryoballoon ablation: back to the drawing board
Heart Rhythm
2010
7
191
2

13

Kubala
M
Hermida
JS
Nadji
G
Quenum
S
Traulle
S
Jarry
G
Normal pulmonary veins anatomy is associated with better AF-free survival after cryoablation as compared to atypical anatomy with common left pulmonary vein
Pacing Clin Electrophysiol
2011
34
837
43

14

Sorgente
A
Chierchia
GB
de Asmundis
C
Sarkozy
A
Namdar
M
Capulzini
L
et al.
Pulmonary vein ostium shape and orientation as possible predictors of occlusion in patients with drug-refractory paroxysmal AF undergoing cryoballoon ablation
Europace
2011
13
205
12

15

Providência
R
Marijon
E
Albenque
JP
Combes
S
Combes
N
Jourda
F
et al.
Rivaroxaban and Dabigatran in patients undergoing catheter ablation of atrial fibrillation
Europace
2014
16
1137
44

16

Siklódy
CH
Minners
J
Allgeier
M
Allgeier
HJ
Jander
N
Keyl
C
et al.
Pressure-guided cryoballoon isolation of the pulmonary veins for the treatment of paroxysmal atrial fibrillation
J Cardiovasc Electrophysiol
2010
21
120
5

17

Sorgente
A
Chierchia
GB
de Asmundis
C
Sarkozy
A
Capulzini
L
Brugada
P
Complications of atrial fibrillation ablation: when prevention is better than cure
Europace
2011
13
1526
32

18

Reddy
VY
Shah
D
Kautzner
J
Schmidt
B
Saoudi
N
Herrera
C
et al.
The relationship between contact force and clinical outcome during radiofrequency catheter ablation of atrial fibrillation in the Toccata study
Heart Rhythm
2012
9
1789
95

19

Marijon
E
Fazaa
S
Narayanan
K
Guy-Moyat
B
Bouzeman
A
Providencia
R
et al.
Real-time contact force sensing for pulmonary vein isolation in the setting of paroxysmal atrial fibrillation: procedural and 1-year results
J Cardiovasc Electrophysiol
2014
25
130
7

20

Stabile
G
Solimene
F
Calò
L
Anselmino
M
Castro
A
Pratola
C
et al.
Catheter-tissue contact force for pulmonary veins isolation: a pilot multicentre study on effect on procedure and fluoroscopy time
Europace
2014
16
335
40

21

Neuzil
P
Reddy
VY
Kautzner
J
Petru
J
Wichterle
D
Shah
D
et al.
Electrical reconnection following PVI is contingent on contact force during initial treatment—results from the EFFICAS I study
Circ Arrhythm Electrophysiol
2013
6
327
33

22

Tilz
RR
Makimoto
H
Lin
T
Rillig
A
Metzner
A
Mathew
S
et al.
In vivo left-ventricular contact force analysis: comparison of antegrade transseptal with retrograde transaortic mapping strategies and correlation of impedance and electrical amplitude with contact force
Europace
2014
16
1387
95

23

Fürnkranz
A
Bordignon
S
Schmidt
B
Gunawardene
M
Schulte-Hahn
B
Urban
V
et al.
Improved procedural efficacy of pulmonary vein isolation using the novel second-generation cryoballoon
J Cardiovasc Electrophysiol
2013
24
492
7

24

Martins
RP
Hamon
D
Césari
O
Behaghel
A
Behar
N
Sellal
JM
et al.
Safety and efficacy of a second-generation cryoballoon in the ablation of paroxysmal atrial fibrillation
Heart Rhythm
2014
11
386
93

25

Chierchia
GB
Di Giovanni
G
Ciconte
G
de Asmundis
C
Conte
G
Sieira-Moret
J
et al.
Second-generation cryoballoon ablation for paroxysmal atrial fibrillation: 1-year follow-up
Europace
2014
16
639
44

26

Andrade
JG
Dubuc
M
Guerra
PG
Landry
E
Coulombe
N
Leduc
H
et al.
Pulmonary vein isolation using a second-generation cryoballoon catheter: a randomized comparison of ablation duration and method of deflation
J Cardiovasc Electrophysiol
2013
24
692
8

27

Chun
KR
Fürnkranz
A
Köster
I
Metzner
A
Tönnis
T
Wohlmuth
P
et al.
Two versus one repeat freeze-thaw cycle(s) after cryoballoon pulmonary vein isolation: the alster extra pilot study
J Cardiovasc Electrophysiol
2012
23
814
9

28

Chierchia
GB
Di Giovanni
G
Sieira-Moret
J
de Asmundis
C
Conte
G
Rodriguez-Mañero
M
et al.
Initial experience of three-minute freeze cycles using the second-generation cryoballoon ablation: acute and short-term procedural outcomes
J Interv Card Electrophysiol
2014
39
145
51

29

Mugnai
G
Chierchia
GB
De Asmundis
C
Sieira-Moret
J
Conte
G
Capulzini
L
et al.
Comparison of pulmonary vein isolation using cryoballoon versus conventional radiofrequency for paroxysmal atrial fibrillation
Am J Cardiol
2014
113
1509
13

30

DeVille
JB
Svinarich
JT
Dan
D
Wickliffe
A
Kantipudi
C
Lim
HW
et al.
Comparison of resource utilization of pulmonary vein isolation: cryoablation versus RF ablation with three-dimensional mapping in the value PVI study
J Invasive Cardiol
2014
26
268
72

31

Schmidt
M
Dorwarth
U
Andresen
D
Brachmann
J
Kuck
KH
Kuniss
M
et al.
Cryoballoon versus RF ablation in paroxysmal atrial fibrillation: results from the German Ablation Registry
J Cardiovasc Electrophysiol
2014
25
1
7

32

Casado-Arroyo
R
Chierchia
GB
Conte
G
Levinstein
M
Sieira
J
Rodriguez-Mañero
M
et al.
Phrenic nerve paralysis during cryoballoon ablation for atrial fibrillation: a comparison between the first- and second-generation balloon
Heart Rhythm
2013
10
1318
24

33

Pérez-Castellano
N
Fernández-Cavazos
R
Moreno
J
Cañadas
V
Conde
A
González-Ferrer
JJ
et al.
The COR trial: a randomized study with continuous rhythm monitoring to compare the efficacy of cryoenergy and radiofrequency for pulmonary vein isolation
Heart Rhythm
2014
11
8
14