A.‐K. Meinshausen, M. Herbster, C. Zwahr, M. Soldera, A. Müller, T. Halle, A.F. Lasagni, J. Bertrand, Aspect ratio of nano/microstructures determines Staphylococcus aureus adhesion on PET and titanium surfaces, Journal of Applied Microbiology, Volume 131, Issue 3, 1 September 2021, Pages 1498–1514, https://doi.org/10.1111/jam.15033
Close - Share Icon Share
Abstract
Joint infections cause premature implant failure. The avoidance of bacterial colonization of implant materials by modification of the material surface is therefore the focus of current research. In this in vitro study the complex interaction of periodic structures on PET and titanium surfaces on the adhesion of Staphylococcus aureus is analysed.
Using direct laser interference patterning as well as roll‐to‐roll hot embossing methods, structured periodic textures of different spatial distance were produced on surfaces and S. aureus were cultured for 24 h on these. The amount of adhering bacteria was quantified using fluorescence microscopy and the local adhesion behaviour was investigated using scanning electron microscopy. For PET structures, minimal bacterial adhesion was identified for an aspect ratio of about 0·02. On titanium structures, S. aureus adhesion was significantly decreased for profile heights of < 200 nm. Our results show a significantly decreased bacterial adhesion for structures with an aspect ratio range of 0·02 to 0·05.
We show that structuring on surfaces can decrease the amount of S. aureus on titanium and PET as common implant materials.
The study highlights the immense potential of applying specific structures to implant materials to prevent implant colonization with pathogen bacteria.
Introduction
Periprosthetic joint infections (PJI) after total joint arthroplasty (TJA) represent a serious complication causing severe suffering, extended hospitalization and challenging revision surgery for the patient. The risk for PJI is about 5·2 % and each revision increases the risk by 10 % to 15 % (Goldberg et al. 1988; Fehring 2000; Smucny et al. 2015). The worst case of PJI results in a sepsis or the need of an amputation (Goldberg et al. 1988; Fehring et al. 2000; Ulrich et al. 2008; Wetters et al. 2013; Smucny et al. 2015; Sousa et al. 2018).
Standard materials used for orthopaedic implants are titanium‐ or cobalt‐base metal alloys, alumina ceramics and polymers. Ultrahigh molecular weight polyethylene (UHMWPE ISO 5834‐3:2019) is commonly used as soft bearing component, for the manufacturing of hip inlays for the acetabular cups or as meniscus component in knee implants. UHMWPE exhibits a low density, sufficient impact strength, good biocompatibility and is at the same time inexpensive and easy to machine (Fisher and Dowson 1991; Sutula et al. 1995). Polyester polyethylene terephthalate (PET) fibres are also used for the production of artificial tendon and ligament replacements. The semi‐crystalline polymer can fix contracting muscles to bone anchor (Beach et al. 1991; Kvist et al. 2005). These fibres exhibit an increased hardness, relatively high strength and good stiffness (Gao et al. 2010; Melvin et al. 2010). Metal alloys, such as pure titanium (ISO 5832‐2), titanium‐base alloys (ISO 5832‐3/‐11) or cobalt‐base alloys (ISO 5832‐4/‐12), are used for the manufacturing of load‐bearing implant components. Titanium implants are favoured for osseointegration since they ensure an ideal bone apposition on cementless femoral stems or acetabular shells (Gibon et al. 2017).
Bacteria are able to colonize all kinds of different implant surfaces in an exogenous or endogenous way (Briandet et al. 2001). The risk of developing a PJI is increased in knee (2 %) in contrast with hip arthroplasty (~1 %) (Andrej Trampuz and Werner Zimmerli 2008). Common pathogens causing a PJI are bacteria from the staphylococcal genus, especially S. aureus and S. epidermidis (Zimmerli and Sendi 2011; Barrett and Atkins 2014; Nair et al. 2014). Interestingly, the microbiological spectrum varies depending on the localization of the joint infection (Marculescu and Cantey 2008; Peel et al. 2012; Tande and Patel 2014). For example, Propionibacterium acnes is found more frequently in shoulder endoprostheses than in other orthopedic implants. (Piper et al. 2009). Anaerobic bacteria are more likely to be present in hip PJI than in knee PJI (Marculescu and Cantey 2008). Therefore, different structures might be needed to achieve the bacteria repellent effect for the broad variety of different species with their different shape and size (Jaggessar et al. 2017). After the adhesion of the bacteria to the implant material, the bacteria are capable of forming a biofilm (Costerton et al. 1999; Webb et al. 2003; Hall‐Stoodley et al. 2004; Høiby et al. 2010; Høiby et al. 2015). The biofilm formed by bacteria consists of a dense network of polysaccharides, in which extracellular DNA, proteins and bacteria are embedded. The biofilm protects the bacteria from external influences, leading to a higher tolerance towards antibiotics and the immune system (Briandet et al. 2001; Flemming and Wingender 2010; Boháčová et al. 2019). There are two basic strategies to prevent biofilm formation, either by inhibition of the biofilm formation or by repelling of bacteria from the start. Each patient undergoing a TJA surgery is often preoperatively treated with a combination of different antibiotics to prevent the perioperative infection (Pavel et al. 1974; Pollard et al. 1979; Mauerhan et al. 1994; Meehan et al. 2009). However, this therapy only prevents the immediate infection directly after surgery. Late infections and haematogenic infections cannot be prevented with this prophylactic therapy (Lentino 2003). Furthermore, the increasing presence of bacterial antibiotic resistance increases the risk for PJI as not all pathogens might be killed with treatment before the surgery (Lentino 2003).
Interestingly, microbial repellency can be achieved by applying a specific surface structure on implant surfaces (Otto 2008; Pavithra and Doble 2008). Various studies investigated the influence of the surface roughness on the adhesion of different bacteria (Esposito et al. 2007; Duarte et al. 2009; Ivanova et al. 2010; Truong et al. 2010; Wu et al. 2011b). A study showed that holes or craters larger than the bacteria in the surface seem to inhibit the formation of bacteria colonies (Epperlein et al. 2017; Zwahr et al. 2019a). Since colony clustering is an essential precondition for the formation of a biofilm, these structures might inhibit biofilm formation. It is known that rough surfaces in the macro and micro scale, however, increase microbial adhesion. In particular, surface profiles in the size of the bacteria (about 1 µm) provide maximum contact area and thereby enhance adhesion. Slightly smaller structure patterns within the micrometer and submicron range decrease the bacterial adhesion (Medilanski et al. 2002; Whitehead et al. 2005; Helbig et al. 2016; Epperlein et al. 2017; Lutey et al. 2018; Zwahr et al. 2019a). In this context, recent studies investigated bacterial adhesion on nanorough titanium surfaces by applying different deposition techniques (Jaggessar et al. 2017), including magneton sputtering (Whitehead et al. 2005; Ivanova et al. 2010) physical vapour deposition (Lüdecke et al. 2013), supersonic cluster beam deposition (Singh et al. 2013) or laser‐based structuring techniques (Cunha et al. 2016; Helbig et al. 2016; Rosenkranz et al. 2016; Epperlein et al. 2017; Lutey et al. 2018; Schwibbert et al. 2019; Florian et al. 2020; Jalil et al. 2020). These studies proved a significant reduction in bacterial adhesion at a mean roughness of 2 to 20 nm. However, the different structuring techniques vary in their uniformity and generate different surface properties regarding porosity, chemical composition (e.g. metal oxides, chemical residues) and surface energy. Since these characteristics influence bacterial adhesion in combination with surface topography, a comparison of results from different techniques is difficult. In addition, the specific geometric parameters including the spatial and vertical surface parameters cannot be applied in a reproducible manner.
Direct laser interference patterning (DLIP) is a fast and cost‐efficient processing technology to generate high‐resolution structured surface features of different morphology. For example, using two laser beams which are overlapped under a certain angle, it is possible to generate an interference pattern with a line‐like periodic intensity distribution (Lasagni et al. 2009; Rodriguez et al. 2009; Bremus‐Koebberling et al. 2012; Oliveira et al. 2012; Zwahr et al. 2019a). Later, based on photothermal or photochemical interaction of the laser light with the material, wave‐like patterns of high homogeneity are formed since the material surface is locally ablated or modified at the interference maxima positions. Furthermore, the DLIP method can be used for treating large‐area hot‐embossing tools, which permit to transfer the surface topography of the metallic tools to polymer foils at processing speeds of several square meters per minute (Soldera et al. 2020).
The present study investigates the potential of DLIP generated structures for inhibition of bacterial adhesion. Therefore, the resulting morphology and roughness of directly treated titanium (Ti) surfaces as well as indirectly processed polyethylene terephthalate (PET) foils using roll‐to‐roll (R2R) hot‐embossing is analysed with confocal microscopy. The adhesion of S. aureus on the different structures in comparison to polished reference surfaces is evaluated. Scanning electron microscopy is applied to visualize the bacterial adhesion. The aim is to identify critical structuring parameters that exhibit the best S. aureus repellence.
Materials and methods
Sample fabrication
Direct laser interference patterning (DLIP) was applied to generate structures on commercially pure titanium discs (Ti grade 4 ISO 5832‐2) with 25 mm in diameter. By varying the angle between the two interfering laser beams, the distribution of the interference intensity was modified and different spatial periods of the periodic wave‐like structures were generated. In addition, the profile height was customized by adjusting the processing speed and laser energy density, also known as laser fluence. A brief description of the DLIP method with schematic illustration is included as supplementary information (Fig. A1(a)). In total, five structures were generated with different spatial periods ranging from 700 nm to 5 µm and profile heights of 100 nm to 800 nm as summarized in Table 1. As reference, a polished Ti‐6% Al‐4%V wrought alloy sample (ISO 5832‐3) with a diameter of 13 mm was used. Further information about this method has been published elsewhere (Voisiat et al. 2019).
Processing and geometric parameters of the line‐like structured Ti samples using DLIP
| Ti sample ID . | Fluence (J cm−2) . | Spatial period (µm) . | Profile height (nm) . |
|---|---|---|---|
| A | 0·9 | ~ 5 | ~ 800 |
| B | 0·29 | ~ 2·2 | ~ 500 |
| C | 0·06 | ~ 1·3 | ~ 200 |
| D | 0·06 | ~ 0·7 | ~ 180 |
| E | 0·47 | No periodicity, random structures | ~ 100 |
| Control | No | No | ~ 200 |
| Ti sample ID . | Fluence (J cm−2) . | Spatial period (µm) . | Profile height (nm) . |
|---|---|---|---|
| A | 0·9 | ~ 5 | ~ 800 |
| B | 0·29 | ~ 2·2 | ~ 500 |
| C | 0·06 | ~ 1·3 | ~ 200 |
| D | 0·06 | ~ 0·7 | ~ 180 |
| E | 0·47 | No periodicity, random structures | ~ 100 |
| Control | No | No | ~ 200 |
Processing and geometric parameters of the line‐like structured Ti samples using DLIP
| Ti sample ID . | Fluence (J cm−2) . | Spatial period (µm) . | Profile height (nm) . |
|---|---|---|---|
| A | 0·9 | ~ 5 | ~ 800 |
| B | 0·29 | ~ 2·2 | ~ 500 |
| C | 0·06 | ~ 1·3 | ~ 200 |
| D | 0·06 | ~ 0·7 | ~ 180 |
| E | 0·47 | No periodicity, random structures | ~ 100 |
| Control | No | No | ~ 200 |
| Ti sample ID . | Fluence (J cm−2) . | Spatial period (µm) . | Profile height (nm) . |
|---|---|---|---|
| A | 0·9 | ~ 5 | ~ 800 |
| B | 0·29 | ~ 2·2 | ~ 500 |
| C | 0·06 | ~ 1·3 | ~ 200 |
| D | 0·06 | ~ 0·7 | ~ 180 |
| E | 0·47 | No periodicity, random structures | ~ 100 |
| Control | No | No | ~ 200 |
For processing the PET foils, the following approach was used. First, nickel sleeves with a thickness of 200 µm, 300 mm wide and with a diameter of 300 mm (Saechsische Walzengravur GmbH, Frankenberg, Germany) were used for DLIP structuring. The initial surface roughness Sa was 61 nm. Prior to the laser process, the sleeves were cleaned from contaminations using ethanol.
Next, the nickel sleeves were processed using a 3D‐DLIP unit employing a rotary axis system (developed at the TU‐Dresden, Germany). The 3D‐DLIP structuring system utilizes a solid‐state ps‐laser with 1064 nm laser wavelength (Edgewave PX200, Germany) and a pulse duration of 10 ps. Line‐like structures with a spatial periods of 4·3 µm, 3·8 µm, 2·5 µm and 1·5 µm were produced. The spot diameter was 190 µm and the employed single pulse laser fluence was 0·36 J cm−2 for all spatial periods. By changing the pulse to pulse overlap, the depth of the periodic structures was varied. The cumulated used fluence (resulting from the different pulse to pulse overlaps) for processing the Ni‐sleeve is listed in Table 2. Finally, the PET film samples were processed using the DLIP treated Ni‐sleeve in a roll‐to‐roll (R2R) hot‐embossing system (R2R Basecoater BC51, Coatema Coating Machinery GmbH, Germany) at an embossing temperature of approximately 90°C and at varying web speeds between 1 m min−1 and 12 m min−1 (Rank et al. 2017). The R2R method is illustrated schematically in the supplementary Fig. A1(b).
Processing and geometric parameters of the line‐like structured PET samples
| PET sample ID . | Used cumulated fluence for Ni‐sleeve (J cm−2) . | R2R web speed (m min−1) . | Spatial period (µm) . | Profile height (nm) . |
|---|---|---|---|---|
| 1 | 7·2 | 2 | 4·3 | ~ 400 |
| 2 | 3·6 | 2 | ~ 100 | |
| 3 | 1·8 | 2 | < 50 | |
| 4 | 7·2 | 2 | 3·8 | ~ 500 |
| 5 | 3·6 | 2 | ~ 200 | |
| 6 | 1·8 | 2 | ~ 100 | |
| 7 | 7·2 | 12 | 2·5 | ~ 600 |
| 8 | 7·2 | 4 | ~ 550 | |
| 9 | 7·2 | 1 | ~ 400 | |
| 10 | 7·2 | 2 | ~ 400 | |
| 11 | 3·6 | 2 | ~ 250 | |
| 12 | 1·8 | 2 | ~ 100 | |
| 13 | 7·2 | 2 | 1·5 | ~ 400 |
| 14 | 3·6 | 2 | ~ 250 | |
| 15 | 1·8 | 2 | ~ 200 | |
| Control | No | No | No | ~ 100 |
| PET sample ID . | Used cumulated fluence for Ni‐sleeve (J cm−2) . | R2R web speed (m min−1) . | Spatial period (µm) . | Profile height (nm) . |
|---|---|---|---|---|
| 1 | 7·2 | 2 | 4·3 | ~ 400 |
| 2 | 3·6 | 2 | ~ 100 | |
| 3 | 1·8 | 2 | < 50 | |
| 4 | 7·2 | 2 | 3·8 | ~ 500 |
| 5 | 3·6 | 2 | ~ 200 | |
| 6 | 1·8 | 2 | ~ 100 | |
| 7 | 7·2 | 12 | 2·5 | ~ 600 |
| 8 | 7·2 | 4 | ~ 550 | |
| 9 | 7·2 | 1 | ~ 400 | |
| 10 | 7·2 | 2 | ~ 400 | |
| 11 | 3·6 | 2 | ~ 250 | |
| 12 | 1·8 | 2 | ~ 100 | |
| 13 | 7·2 | 2 | 1·5 | ~ 400 |
| 14 | 3·6 | 2 | ~ 250 | |
| 15 | 1·8 | 2 | ~ 200 | |
| Control | No | No | No | ~ 100 |
Processing and geometric parameters of the line‐like structured PET samples
| PET sample ID . | Used cumulated fluence for Ni‐sleeve (J cm−2) . | R2R web speed (m min−1) . | Spatial period (µm) . | Profile height (nm) . |
|---|---|---|---|---|
| 1 | 7·2 | 2 | 4·3 | ~ 400 |
| 2 | 3·6 | 2 | ~ 100 | |
| 3 | 1·8 | 2 | < 50 | |
| 4 | 7·2 | 2 | 3·8 | ~ 500 |
| 5 | 3·6 | 2 | ~ 200 | |
| 6 | 1·8 | 2 | ~ 100 | |
| 7 | 7·2 | 12 | 2·5 | ~ 600 |
| 8 | 7·2 | 4 | ~ 550 | |
| 9 | 7·2 | 1 | ~ 400 | |
| 10 | 7·2 | 2 | ~ 400 | |
| 11 | 3·6 | 2 | ~ 250 | |
| 12 | 1·8 | 2 | ~ 100 | |
| 13 | 7·2 | 2 | 1·5 | ~ 400 |
| 14 | 3·6 | 2 | ~ 250 | |
| 15 | 1·8 | 2 | ~ 200 | |
| Control | No | No | No | ~ 100 |
| PET sample ID . | Used cumulated fluence for Ni‐sleeve (J cm−2) . | R2R web speed (m min−1) . | Spatial period (µm) . | Profile height (nm) . |
|---|---|---|---|---|
| 1 | 7·2 | 2 | 4·3 | ~ 400 |
| 2 | 3·6 | 2 | ~ 100 | |
| 3 | 1·8 | 2 | < 50 | |
| 4 | 7·2 | 2 | 3·8 | ~ 500 |
| 5 | 3·6 | 2 | ~ 200 | |
| 6 | 1·8 | 2 | ~ 100 | |
| 7 | 7·2 | 12 | 2·5 | ~ 600 |
| 8 | 7·2 | 4 | ~ 550 | |
| 9 | 7·2 | 1 | ~ 400 | |
| 10 | 7·2 | 2 | ~ 400 | |
| 11 | 3·6 | 2 | ~ 250 | |
| 12 | 1·8 | 2 | ~ 100 | |
| 13 | 7·2 | 2 | 1·5 | ~ 400 |
| 14 | 3·6 | 2 | ~ 250 | |
| 15 | 1·8 | 2 | ~ 200 | |
| Control | No | No | No | ~ 100 |
Four different spatial period groups were produced in this case. For each periodic group different profile heights were obtained by either varying the pulse to pulse overlap during the fabrication of the Ni‐sleeve or by changing web speed in the R2R process, as listed in Table 2. A total of 15 samples with structure heights in the range of 30 nm to 600 nm were generated and tested (Table 2). The PET reference was the untreated source material. Each specimen was customized cut to a size of 8 × 8 mm2.
Surface morphologies and roughness
3D topographic images were obtained with a confocal microscope (μsurf expert, NanoFocus AG, Oberhausen, Germany). The measuring field of 163 × 163 µ m2 was acquired with a 100× magnification objective and a vertical resolution of 1 nm. All samples were measured at three different locations. Roughness parameters, including the arithmetic average roughness (Ra), the average maximum profile height (Rz), the core groove depth (Rk DIN EN ISO 13565‐2) and mean line spacing of roughness peaks (RSm), were determined according to the standard ISO 4287. Representative 3D surface images and 2D roughness profiles were obtained on an extracted area of 20 × 20 µm2.
Bacterial adhesion test
To test the bacterial adhesion of the structured PET and Ti surfaces, the samples were incubated three times with S. aureus SH1000 pSB2035 (Qazi et al. 2001). This S. aureus strain constitutively expresses the green fluorescent protein (GFP). S. aureus was cultured overnight in tryptic Soy Broth (TSB). The structured test plates were incubated with 1 ml TSB media and inoculated with a 1:1000 dilution of the S. aureus overnight culture that was adjusted to an optical density (OD600 nm) of 1 for 24 h at 37°C. To prevent contamination, we inoculated the overnight culture from our glycerol stock solution and worked under standard sterile conditions. The metal disks were sterilized using heat inactivation. The medium was removed and the PET and Ti plates were washed twice with phosphate buffered saline (PBS) and subsequently fixed with 4 % paraformaldehyde for 30 min at room temperature. The amount of adhering bacteria was quantified in 630× magnification microscopic pictures (Zeiss fluorescence microscope Axio Observer.Z1, Jena, Germany). The fluorescence of the GFP was excited by a wavelength of 488 nm and a read out of 500 – 550 nm. Pictures from five different randomly chosen, representative areas from each sample were taken and evaluated by counting the amount of bacteria with ImageJ software. In ImageJ we set a threshold to reduce the background fluorescence and to measure only the intensity of the GFP fluorescence (Bui et al. 2020).
Cleaning and sterilization of the samples
Before in vitro incubation, all PET samples were cleaned in pure isopropanol and dried under sterile conditions. Afterwards they were placed under UV light for 24 h. The Ti specimens were cleaned by incubating them for 30 min in 10 ml of RIPA Buffer (150 mmol l−1 sodium chlorid, 1·0 % Triton X, 0·5 % sodium deoxycholate, 0·1 % SDS, 50 mmol l−1 Tris) at room temperature. RIPA is a lysis buffer used for cell lysis and protein solubilization. This reagent is capable of extracting cytoplasmic, nuclear and membrane proteins. Afterwards the platelets were incubated with Trypsin for 30 min at 37°C under sonification. The Ti specimens were rinsed with distilled water and incubated in 70 % ethanol. For sterilization, the Ti plates were flamed for sterilization and placed under UV light for 24 h. Applying this cleaning procedure, the Ti specimens were reused three times. After cleaning, all plates were checked for residual fluorescence indicating adhering bacteria by microscopy. No evidence of GFP fluorescence was observed on the Ti surfaces after the extensive cleaning process using RIPA buffer and trypsin under sonication to remove remaining proteins and adhering bacteria. The PET surfaces were not reused, but new samples were used for each experiment.
Scanning Electron Microscopy analysis
Before or after incubation with S. aureus, all PET and Ti surfaces were analysed using scanning electron microscopy (SEM, FEI Scios DualBeam, Thermo Fisher Scientific, Waltham, USA). SEM investigations were conducted using an acceleration voltage of 10 – 15 keV. Regions of interest were identified using an imaging mode at lower magnifications (2000×). Surfaces were imaged with a secondary electrons (SE) detector.
Statistical analysis
If not stated otherwise, the data are presented as mean ± standard deviation (SD). The Friedman test was applied to test for statistical significance. The level of significance was set at P < 0·05 for all statistical tests. Statistical analysis was performed using GraphPad Prism (ver. 7, GraphPad Software, San Diego, USA) and Origin software 2018 (OriginLab Corporation, Northampton, USA). The graphs presented in this study show the mean values with standard error of mean (SEM).
Results
Characterization of structures
Different periodic profile distances with varying profile heights were generated by DLIP and R2R hot‐embossing. Figure 1 shows representative 3D confocal microscopic images of the investigated PET and Ti surfaces. In general, the surfaces exhibit a high regularity and homogeneity of the line‐like structures. The shapes of all structures are similar and independent of the decrease of surface roughness. We detected only few surface defects on Ti samples that exhibit spatial periods smaller than 1 µm. These defects deteriorate the periodicity of the structures leading to a reduced homogeneity (Fig. 1c,d lower row). Superimposed on the DLIP microstructures, smaller quasi‐periodic features were also visible on both the Ti and PET samples that could be identified as the so‐called laser‐induced periodic surface structures (LIPSS). Figure A2 of the supplementary information shows exemplary SEM images of structured (a) Ti and (b) PET indicating the polarization of the laser beams relative to the grooves’ alignment. In both Ti and PET micrographs, low spatial frequency LIPSS (LSFL) and high spatial frequency LIPSS (HSFL) are indicated. It is worth to notice that the LIPSS observed on the PET surface, were produced originally on the Ni sleeve and transferred to the polymer during the hot embossing process. In both materials the LSFL were aligned perpendicular to the radiation polarization and had a spatial period of approximately 800 nm, which is ~ 80 % of the used wavelength. These characteristics are in agreement with previous works on laser irradiation on metals with ultra‐short pulses in the near‐infrared spectrum (Römer et al. 2009; Vincenc Obona et al. 2011). It is commonly accepted that LSFL are produced by the interaction of the laser radiation with an electromagnetic wave scattered at the rough surface of strongly absorbing materials, such as metals, involving the excitation of surface plasmon polaritons (Bonse et al. 2017). In turn, the HSFL indicated in the SEM images of the supplementary Figure A2 presented a spatial period of approximately 150 nm and were aligned parallel to the polarization. Although the origin of the HSFL is still under debate, several studies confirm the presence of such nanostructures on Ti and Ni upon irradiating the surfaces with ultra‐short pulses as observed in our work (Wu et al. 2011a; Höhm et al. 2015; Lang et al. 2017; Zwahr et al. 2019b).
Morphological investigation of the tested samples with different structures. (a) 3D confocal microscopic images of the PET samples of varying heights and spatial periods of 4·3 µm, 3·8 µm, 2·5 µm and 1·5 µm. (b) Roughness profiles over a length of 20 µm perpendicular to the line structures of the investigated PET surfaces. (c) Exemplary images of the Ti samples showing different linear profile heights and spatial periods of 5 µm, 2·5 µm, 1 µm and 0·5 µm. (d) Roughness profiles of the investigated Ti samples.
The determined roughness parameters are summarized in Fig. 2. The mean values of the mean line spacing of profile peaks (RSm) vary in the range of 1·46 µm to 5·1 µm for PET samples and between 0·69 µm to 5·18 µm for Ti surfaces (Fig. 2a). Based on the profile distances, four PET gradation groups were generated with mean values of 4·3 µm (sample number: 1–3), 3·8 µm (4–6), 2·5 µm (7–12) and 1·5 µm (13–15). The corresponding mean values of the average profile heights (Rz) reveal variations in the size range of 45 nm to 550 nm for the PET specimen. The aspect ratio is defined in this work as the ratio between Rz and RSm (Fig. 2c). Accordingly, each periodic PET group exhibit different aspect ratios, as illustrated in Fig. 2d. The corresponding mean values of the arithmetic mean roughness (Ra) varied between 4 nm to 143 nm for the PET specimens. The Ti samples exhibit mean values of Rz in the range of 162 nm to 772 nm with corresponding Ra values between 25 nm and 162 nm. The reference sample of polished TiAl6V4 alloy exhibits the lowest mean Ra value of 9·6 ± 2 nm. Figure 2d demonstrates the sample roughness distribution according to the arithmetic mean roughness value (Ra) and the corresponding aspect ratio. The mean Ra values range between 8 nm to 170 nm with different aspect ratios (range 0·01 to 0·35).
Results of the confocal roughness measurements. (a) Mean values ± SD of the mean line spacing of profile peaks (RSm) that represent the spatial distances of structures on the PET and Ti samples. (b) Mean values of the average maximum profile height (Rz) of the PET and Ti structures. (c) Correlation of the mean Rz and RSm values of the PET and Ti structures. (d) The arithmetic mean roughness Ra of the structured surfaces and the corresponding aspect ratio (Rz/RSm) of the produced samples. (
) Ti and (
) PET.
Influence of structures on bacterial adhesion on PET and Ti surfaces
The in vitro quantification of S. aureus adhesion to the structured PET surfaces is shown and summarized in Fig. 3. An unstructured PET surface was used as control surface (mean value of adherent bacteria 2512·0 ± 423 on an area of 283·48 µm × 179·54 µm). The amount of bacteria adhering to the surfaces did differ significantly depending on the roughness height and the spatial distance of the sample. The sample 2 (1146 ± 273 P = 0·0038), sample 6 (1131 ± 958 P = 0·0175) and sample 11 (1065 ± 668 P = 0·0411) showed significantly less adherent bacteria compared to the control group. In contrast, we observed higher number of bacteria compared to the control group in sample 4 (2586 ± 760) with a spatial distance of 3·8 µm. Also the samples 7 (2234 ± 1579) and sample 8 (2234 ± 745) with a spatial period of 2·5 µm as well as sample 14 (2391 ± 1775) with lowest spatial period of 1·5 µm seemed to increase the number of bacteria adhering. A selection of representative microscopic pictures is given in Fig. 3b. As the S. aureus strain produced GFP the green area represents the amount of adhering bacteria.
In vitro testing of antibacterial properties of structured PET surfaces. (a) Quantitative analysation of the number of Staphylococcus aureus on the PET surfaces. The bacteria were incubated on the surfaces for 24 h and then fixed for microscopy, representative areas (n = 4) were microscoped and the number of bacteria were analysed via ImageJ Tool. The diagram shows the mean values with SEM. The fluorescence images of the bars marked in light grey can be seen in b, the numbering of the bar charts from 1 to 6 indicates the order of the fluorescence figures in b. The surfaces marked with an asterisk were subsequently examined with SEM (Fig. 4). (b) The fluorescence microscopy images were taken at a magnification of 630×. The scale bars (white bars) are 50 µm. *P < 0·05; n, total number of measurements, N, number of bacteria.
The location of S. aureus attaching to the structured PET samples was further investigated using SEM. Representative electron micrographs of bacteria adhering to structures with a spatial periods in the range of 2·5 µm are shown in Fig. 4. We observed different bacteria adhesion patterns with decreasing profile heights. Roughness profile heights of about 550 nm, representing half of the bacteria size, lead to a linear orientation of single or coupled bacteria along the parallel profile grooves (Fig. 4a). In contrast, a structure with decreased profile height in the range of 400 nm (Fig. 4b) exhibited an increased amount of bacteria within the profile grooves of the structure. Clusters of at least four bacteria accumulated inside the grooves. These bacterial accumulations were present within all grooves, covering several machining lines. In the texture with a profile height of about 250 nm only few single bacteria adhered to the surface. The present bacteria did only vaguely correlate with the groove patterning. Furthermore, they appear to adhere only with limited superficial contact without close contact to each other (Fig. 4c). These results are in line with the quantification data.
Electron micrographs showing the location of Staphylococcus aureus on structured PET samples of a similar spatial distance (RSm of 2·5 µm) with different height profiles (Rzrange of 550 nm to 250 nm). Different localization spots of bacteria on the structures were observed (SEM magnification top: 6500×, down: 20 000×). (a) Coupling of S. aureus and linear orientation within the profile grooves. (b) Grouping of several bacteria in the roughness depths and connecting over profile peaks. (c) Single bacteria that adhere irregularly to the shallow structures.
The results of adhering S. aureus on structured Ti surfaces are shown in Fig. 5. The quantification of adherent bacteria is given in Fig. 5a. As control surface (4135 ± 98 bacteria), a polished TiAl6V4 wrought alloy (Ra = 9·6 ± 2·1 nm) sample was used. In comparison to the unstructured control sample all structured surfaces show a significant reduction in the number of adhering bacteria. We found that with decreasing profile height from 1·5 µm (sample A) to 180 nm (sample D) the number of bacteria was decreased more efficiently. At the lowest profile height (1648 ± 140 bacteria; P = 0·0376) the strongest reduction was observed. However, no significant differences were measured between the different applied structures (A–E). Fig. 5b shows representative microscopic images that were used for the quantification of the bacteria on the tested structured samples. S. aureus bacteria are depicted as green fluorescence.
In vitro testing of antibacterial properties of structured Ti surfaces. (a) Quantitative analysis of the number of Staphylococcus aureus on the Ti surfaces. The bacteria were incubated on the surfaces for 24 h and then fixed for microscopy, representative areas (n = 4) were microscoped and the number of bacteria were analysed via ImageJ Tool. Shown is the mean with SEM. The numbering of the bar charts from 1 to 6 indicates the order of the fluorescence figures in b. The surfaces marked with an asterisk were subsequently examined with SEM (Fig. 6) (b) The fluorescence microscopy images were taken at a magnification of 630×. The scale bars (white bars) are 50 µm. *P < 0·05; n, total number of measurements, N, number of bacteria.
In order to visualize the localization of S. aureus on the different structured Ti surfaces, SEM was applied. At a magnification of 6500×, the linear structures on the Ti surface are present as parallel filament‐like structures that vary in thickness and length with few minor defects and pores (Fig. 6, upper row). At higher magnification (20 000×) S. aureus bacteria can be seen roundly shaped and adhering in grape‐like clusters. Interestingly, fewer clusters are observed with a decreased profile height (Fig. 6b,c). These results are in line with the quantification data.
Representative electron micrographs of the location of adhering Staphylococcus aureus on the structured Ti surfaces. Along and within the porous linear profile lines, different bacterial clusters are observed (SEM magnification top: 6500×, bottom: 20 000×). (a) Multiple bacteria clusters cover the structured surface. The bacteria are arranged along the parallel filament lines. (b) Structure with a spatial distance pattern of 2·2 µm shows adhering bacteria clusters. (c) Ti surface with very fine structures in the height of 180 nm show very few bacteria clusters.
Discussion
The aim of this study was to investigate the effect of periodic structures of wave‐shaped morphology with varying surface nanoroughness on the adhesion of S. aureus. All structured samples were compared with clinical relevant control surfaces regarding the short‐term adhesion of S. aureus.
Different repellency strategies for medical implants can be applied in order to generate antibacterial properties either by film deposition, smoothening or roughening of the machined surfaces (Esposito et al. 2007; Jaggessar et al. 2017). All surface modifications are supposed to hinder the adhesion of bacteria to the implant surface and inhibit the formation of a biofilm (Hasan et al. 2013). Besides biochemically modified surfaces, there is also the possibility of surface structuring. Previous studies have already shown that the surface roughness has an influence on the adhesion of micro‐organisms (Esposito et al. 2007; Duarte et al. 2009). Rougher surfaces tend to increase the microbial adhesion due to a greater surface area, mechanical provision of protection and the presence of grooves and pits, which influence positively the shear forces of the pathogens (Nyvad and Fejerskov 1987; Morgan and Wilson 2001). Different techniques exist to apply desired structures to implant surfaces. However, low process costs, high pattern resolution and complexity for nanoscale features are the current technological challenges (Lunelli et al. 2012). The present study applied DLIP and R2R hot‐embossing methods to generate periodic structures on PET and Ti samples. The characterization of the produced structures demonstrated a high homogeneity, consistency and reproducibility of the wave‐like profiles for all tested profile heights on PET. In contrast, the Ti samples exhibited variable filament‐like structural surfaces with inhomogeneities being present in structures with spatial distances smaller than 1 µm (Figs 1 and 6). These differences in the applied structure contour can be attributed to the laser treatment that transfers energy to the material surface. In particular, it is known that when using ps‐laser sources, LIPSS can be formed which cover the surface of the periodic DLIP structures (Fu et al. 2019). Additionally, during this process the increasing temperatures melt the surface of the titanium bulk material, change the surface constitution and induce the formation of thicker oxide layers that is accompanied by microporosity. The DLIP generated thin reactive layers on Ti have been previously investigated by Zwahr et al. (2019a). Cunha et al. (2016) studied the effects of surface characteristic of laser structured Ti surfaces on the adhesion of bacteria. The study verified that the chemical composition of the oxide layer and surface wettability have a minor effect on S. aureus adhesion in contrast to surface topography that was also confirmed by Helbig et al. (2016). Further research regarding the DLIP processing parameters is required in order to increase the structural resolution of the Ti structures with minimal spatial distance of smaller than 1 µm.
The profile heights of the different patterns vary between 30 nm to 600 nm for PET and 100 nm to 800 nm for Ti and result in mean Ra values of 4 nm to 143 nm and 25 nm to 162 nm for PET and Ti, respectively. This range of applied nanoroughness was chosen because previous studies found the lowest bacterial adhesion in this range (Whitehead et al. 2005; Ulrich et al. 2008; Sousa et al. 2018). In combination with varying spatial distances of the wave‐like shapes, the structures cover an aspect ratio range from 0·01 to 0·35. The in vitro testing results revealed that the vertical and horizontal feature size of the structures influenced the amount and location of adhering bacteria. Based on the results of quantitative in vitro tests, S. aureus adhesion correlated with the aspect ratio of the structures and exhibited a maximum threshold at 0·15. As shown in Fig. 7, minimal bacterial adhesion is present for an aspect ratio in the range from 0·02 to 0·05 followed by few bacterial contacts between 0·05 and 0·12 in comparison to the control. We did not observe a marked difference between Ti and PET structured samples. Nevertheless, a further decrease in aspect ratio (<0·01), similar to surface smoothening, results in an increased microbial adhesion. Different publications verified an increased bacterial adhesion on titanium for a roughness range of Ra between 1 nm to 4 nm which corresponds to polished surfaces (Ivanova et al. 2010; Truong et al. 2010; Wu et al. 2011b). This effect is supported by the fact that very smooth surfaces promote an increased production of extracellular polymeric substances (EPS) which are essential for the biofilm formation (Whitehead and Verran 2007). Interestingly, an aspect ratio of 0·15 shows a similar increased adhesive susceptibility compared to the smooth reference samples. The structures with an aspect ratio of 0·15 cover a wide Ra range of 50 nm to 150 nm, but the results highlight that this characteristic ratio between nanometer sized feature to the spatial period in combination with the superimposed LIPSS enables an ideal surface contact of S. aureus as well as an increased bacteria agglomeration within the forming recesses. The resulting cell‐surface interaction promotes the aggregate formation of S. aureus as perquisite for colonization (Fig. 7a). An aspect ratio that exceeds 0·2 results in a reduction of adhering S. aureus. Comparable adhesive behaviour was observed for S. aureus strains on PET as well as on Ti surfaces. The aspect ratio influences the attachment contact area between microorganism and material and consequently the adhesion and colonization behaviour (Singh et al. 2011). A minimal contact area of the bacterium with the surface reduces the interaction energy and results in single or coupled bacteria that adhere to the surface (Bhushan et al. 2009; Guo et al. 2011; Renner and Weibel 2011). While smooth surfaces provide an ideal large contact area for adhering bacteria, our results highlight that surface profiles with characteristic dense features of nanometer height, smaller than the bacterium size, reduce the contact area for S. aureus due to inhibited penetration of the bacteria into the recesses of the structures. The limited attachment of the bacteria to individual surface features decreases the total contact area and thus reduces the adhesion strength of the individual bacterium to the material in contrast with the polished reference samples. This inhibitory adhesion was also demonstrated by Cunha et al. (2016) who observed a reduction in S. aureus adhesion on femtosecond titanium LIPPS and nanopillars in contrast to polished titanium samples. Similar antimicrobial effects of LIPPS and nanopillars on stainless steel were also demonstrated by Lutey et al. (2018) for E. coli and S. aureus. These identified effects can explain the significant reduction of adhering bacteria in the aspect ratio range of 0·02 to 0·12 on our samples.
Schematic representation of the influence of the aspect ratio on the number of adhering bacteria for Ti and PET. (a) Number of Staphylococcus aureus according to the aspect ratio of the tested structures on PET and Ti specimens (
) PET and (
) Ti. (b) Scheme that illustrates the influence of different aspect ratio of 0·3, 0·2 and 0·1 on bacteria adhesion based on the finding on tested structures on PET and Ti specimens (
) 0·3; (
) 0·2 and (
) 0·1.
Based on current publications, the size and shape of the bacteria play an important role during the colonization of different material surfaces. Epperlein et al. (2017) showed a differentiated adhesion behaviour between the bacterial strains E. coli and S. aureus on different types of steel LIPSS. It can be attributed to the variation in shape and size of rod‐like E. coli and coccoid formed S. aureus as well as on the bacterial EPS production. It has already been shown that organisms that are smaller than the surface patterns exhibit a higher adhesion force due to the increased contact area (Whitehead and Verran 2006; Scardino et al. 2008). This trend was also observed in our study. In case of smaller grooves than bacteria size less adhering S. aureus were detected compared to bigger grooves. As shown in Figures 4 and 6, bacteria preferred to bind directly in the grooves between the periodic structures.
This bacterial orientation maximizes the contact area, which in turn leads to an increased adhesion and colonization of multiple micro‐organisms (Hochbaum and Aizenberg 2010; Perera‐Costa et al. 2014; Vasudevan et al. 2014). It has been described previously that differently scaled topographies can reduce the biofilm formation and the adhesion of pathogens (Scardino et al. 2009; Scardino and de Nys 2011; Bixler and Bhushan 2012; Crawford et al. 2012; Hasan and Chatterjee 2015; Nir and Reches 2016; Elbourne et al. 2017). In this context, Majhi and Mishra (2020) demonstrated that laser patterning of surfaces represents an appropriate technology to reduce bacterial adhesion, since it affects the surface morphology as well as its wettability. Majhi and Mishra (2020) showed that the capability of S. aureus for the binding on a surface is dependent on the surface wettability and the average surface roughness which was tested with nanopillars with a size of 0·8 µm to 1·3 µm. According to Lam et al. (2016) sharp nanostructures are able to break the cell wall of pathogens and can kill them immediately. Anselme et al. (2010) investigated the interaction between bacteria and surface at a nanoscale and found that bacterial appendages (flagella, fimbriae etc.) determine the interaction with nanoscale patterns and thus influence bacterial adhesion. Puckett et al. (2010) compared in an in vitro study the adhesion of S. aureus, S. epidermidis and Pseudomonas aeruginosa on conventional nanosmooth to nanorough titanium produced by electron beam evaporation. A significant reduction for all tested bacteria strains was observed on structured titanium surfaces. Similar to our results Helbig et al. (2016) proved the inhibition of bacterial colonization by sub‐cell sized DLIP topographies using S. epidermidis and E. coli irrespective of the surface chemistry. The results from our study show that the mean roughness for periodic features of a surface is not the main influencing parameter for adhesion of bacteria (Supplementary Figure A3), but rather the aspect ratio of the DLIP structures determines bacterial adhesion preferences.
Depending on the separation distances between the structures, different adhesion interactions occur. While small separation distances (below 5 nm) mainly involve molecular interactions, macroscopic surface properties, such as surface free energy or hydrophobicity, are also important factors for the bacterial adhesion (Busscher and Weerkamp 1987; Cao et al. 2011). Increase of surface roughness within the macro, micro and nano scale enlarges the potential contact area, thereby creating a more effective and attractive adhesive surface for the bacteria (Preedy et al. 2014).
There are some limitations regarding this study. This study is limited to the investigated materials of PET and Ti, line‐like periodic geometries and to only one bacteria strain. As other studies demonstrated previously, the adhesion and biofilm formation of laser‐structured surfaces depend strongly on the type of microorganism tested on the respective surface (Lutey et al. 2018). We are aware of the limitation of our study that only one pathogen was tested on the structured surfaces. However, the most common species found in PJI is Staphylococcus. Therefore, we focused on S. aureus in this work (Arciola et al. 2005a; Arciola et al. 2005b). We are aware that a variety of other pathogens are associated with PJI, such as P. aeruginosa, K. pneumonia and E. coli (Peel et al. 2012). These pathogens differ in size and shape (Young 2007) and therefore might exhibit another attachment profile compared to S. aureus. Further investigations using other pathogens are needed to assess the bacteria repellent effect of the structures in this manuscript in a broader range of PJI related bacteria. In addition, this study did not investigate the effect of the aspect ratio on a potential biofilm formation. However, we assume that the aspect ratio affects the adhesion behaviour of bacteria differently, if the bacteria are cultivated for a longer period due to the extracellular matrix secretion. Future work should consequently include longer cultivation times to investigate the influence of the aspect ratio on the biofilm formation. The structured Ti samples were limited to only four tested structures making it difficult to finally draw conclusions for influence of the tested patterns on the adhesion of bacteria. Future investigations will be needed to analyse a broader variety of structured Ti surfaces and their interaction with S. aureus as well as other common PJI pathogens. Furthermore, our findings regarding the various structures are limited with regard to the applied sterilization process that is different compared to the standard sterilization used by implant manufacturers. These differences might have an impact on the adhesion of bacteria. In addition, the reference sample was a polished TiAl6V4 disc, which represents a very common implant alloy. Future studies are envisaged to apply DLIP on TiAl6V4 alloy and UHMWPE as well as other relevant implant materials used for manufacturing of endoprosthesis.
This study showed that structuring on implant materials critically influences the adhesion and localization of S. aureus on the surface. We found that the aspect ratio was the main critical parameter for bacterial adhesion. A marked decrease of bacterial adhesion was observed for an aspect ratio range of 0·02 and 0·05. These findings will help to develop surface structures on implant materials that exhibit intrinsic bacteria repellent properties.
Acknowledgements
The authors acknowledge the excellent technical assistance of Carolin Schneider, Anja Schröder and Mandy Könnecke.
Conflict of Interest Statement
The authors certify that they have no affiliations with or involvement in any organization or entity with any financial interest (such as honoraria, educational grants or patent‐licensing arrangements), or non‐financial interest (such as personal or professional relationships, affiliations, knowledge or beliefs) in the subject matter or materials discussed in this manuscript.
Author’s contribution
AKM: performed investigation, interpreted data, drafted the manuscript; MH: performed analyses, interpreted data, drafted the manuscript; CZ: helped with material preparation and sample production; MS: sample production, helped with writing the manuscript; AM: provided lab bacteria strain and microbiological expertise; TH: supervised the technical data, interpretation of data; AFL: developed study design, helped with writing the manuscript; JB: developed study design, interpretation of data, helped with writing the manuscript.
Disclaimer
This work was in part conducted within the context of the International Graduate School ABINEP and MEMoRIAL at Otto von Guericke University (OVGU) Magdeburg, Germany, kindly supported by the European Structural and Investment Funds (ESF, 2014–2020) under the program “Sachsen‐Anhalt WISSENSCHAFT Internationalisierung” (project no. ZS/2016/08/80645 and project no. ZS/2016/08/80646). This research and development project was also partially funded by the German Federal Ministry of Education and Research (BMBF) within the “Unternehmen Region, Innovation and structural transformation/WIR!” program, under contract number 03WIR2003B and implemented by the Project Management Agency Jülich (PTJ). M.S. acknowledges the Alexander von Humboldt Foundation. The authors are responsible for the content of this publication. The authors, their immediate families, and any research foundations with which they are affiliated have not received any financial payments or other benefits from any commercial entity related to the subject of this article.
References
Author notes
Ann‐Kathrin Meinshausen, Maria Herbster, Andrés Fabián Lasagni and Jessica Bertrand contributed equally to this work.